background image

Important Information 

— Please Read Before Use

 

 

-2- 

 

For Veterinary use only 

 

 

Signal words

 

The following signal words are used throughout this manual: 

 

: It indicates a potentially hazardous situation, if not be avoided, could result in death 
or serious injury. 

 

: It indicates a potentially hazardous situation, if not be avoided, may result in minor or 
moderate injury. It may also be used to alert against unsafe practices or potential 
equipment damage. 

: It indicates additional helpful information. 

 
 

Warnings and cautions

 

Follow the warnings and cautions below when handling this instrument. Information below is to be 
supplemented in each chapter. 

 

 

After  using  this  endoscope,  clean,  disinfect  and  store  it  according  to  the  instruction  manual. 

Before endoscopy, remove any metallic objects from the patient. 

Do  not  strike,  hit,  or  drop  the  endoscope’s  distal  end,  insertion  tube,  bending  section,  control 
section, universal cord, or endoscope connector. Also, do not bend, twist or pull the 

endoscope’s 

distal  end,  insertion  tube,  bending  section,  control  section,  universal  cord,  or  endoscope 
connector  with excessive force. 

Never perform angulation control forcibly or abruptly. 

Never insert or withdraw the endoscope’s insertion tube while the bending section is locked in 
position. 

Never  operate  the  bending  section,  feed  air  or  perform  suction,  insert  or  withdraw  the 
endoscope’s  insertion  tube,  or  use  endotherapy  accessories  without  viewing  the  endoscopic 
image or while the image is frozen or magnified. 

Do  not  touch  the  endoscope  connector  immediately  after  removing  it  from  the  endoscope 
imaging processor because it is extremely hot. 

When  the  endoscopic  image  does  not  appear  on  the  monitor,  the  sensor  may  have  been 
damaged. Turn the imaging processor OFF immediately. 

If it is difficult to insert the endoscope, do not forcibly insert the endoscope; stop the endoscopy. 

Be  sure  to  inspect  that  the  bending  section  bends  smoothly  by  touching  it  with  hands  before 
inserting the endoscope into the patient. Immediately stop using it and withdraw it from the patient 
if any irregularity is observed. 

Keep  the  insertion  section  and  bending  section  as  straight  as  possible  during  endoscopic 
treatment to give smooth access to the endoscopic instrument. 

 

 

CAUTION

 

NOTE

 

Summary of Contents for VET-1335

Page 1: ...VET 6010 VET 3512 VET 8028HD VET 9833HD VET 9830 VET 9830HD VET 1335 VET 1335HD VET 8015 VET 8015HD VET 9215 VET 9215HD...

Page 2: ...ict Shanghai 201108 P R China Zipcode 201108 Tel 86 21 67681019 Fax 86 21 67681019 Website www aohua com Important information Please Read Before Use Before use thoroughly review this manual Please ke...

Page 3: ...2 2 Performance characteristics 11 2 3 Product characteristics 11 2 4 Specifications 11 2 4 1 Operating environment 11 2 4 2 Interconnected equipment specifications 12 2 4 3 Endoscope specifications 1...

Page 4: ...ing of the video endoscope 24 4 5 5 Rinsing 27 4 5 6 Disinfection of the endoscope 27 4 5 7 Rinsing after disinfection 28 4 5 8 Drying 29 4 5 9 Other accessories cleaning disinfection and sterilizatio...

Page 5: ...dge This device must be operated by a medical practitioner capable of safely performing endoscopy after operation technique training Ancillary Equipment The safety of the endoscope does not only rely...

Page 6: ...control section universal cord or endoscope connector with excessive force Never perform angulation control forcibly or abruptly Never insert or withdraw the endoscope s insertion tube while the bend...

Page 7: ...re etc if the endoscope is not used by following the instruction manual damages or malfunctions may result CAUTION Do not touch the electrical connector inside the electrical connector Do not apply sh...

Page 8: ...and Symbols Type BF equipment Protective earth ground Caution Equipotentiality This way up Refer to instruction manual Date of manufacture Manufacturer Keep dry Use by date Keep away from sunlight Ser...

Page 9: ...Endoscope VET 3512 1 1 Checking the package contents list CAUTION Check all items in the package against the components listed below If any component is missing or damaged do not use the item contact...

Page 10: ...Checking the Package Contents 6 For Veterinary use only Cleaning Brushes Leakage Tester Cleaning Kit Channel Plug Sealing Plug Water Bottle Suction Valve User Manual Biopsy Valve Air water Valve 120 1...

Page 11: ...02 Nomenclature and Functions 7 For Veterinary use only Figure 2 1 1 Endoscope VET 6010 VET 9833HD VET 9830 HD VET 1335 HD VET 8028HD VET 8015 HD VET 9215 HD 2 1 Nomenclature and Functions...

Page 12: ...his cylinder 6 Air water cylinder Attach the air water valve to this cylinder 7 Biopsy Valve This valve is attached to the instrument channel inlet and the endotherapy accessory is inserted or a syrin...

Page 13: ...ansmits light from the processor to the endoscope 17 Suction connector Connects the endoscope to the suction tube of the suction pump 18 S cord connector mount Connects the endoscope with electrosurgi...

Page 14: ...serted 4 Suction valve This valve is depressed to active suction The valve is used to remove fluids from the patient 5 UP DOWN angulation control knob When turn this knob in the U direction the bendin...

Page 15: ...processor of VET OR series Feeding of air and water and suction operation could be activated by operating Air Water valve inapplicable to VET 3512 or suction valve This equipment is classified as type...

Page 16: ...0mm 3000mm 3500mm 600mm 1500mm 1500mm 2 4 3 Endoscope specifications Model VET 6010 VET 8028 HD VET 9833HD VET 9830 HD VET 1335 HD VET 3512 VET 8015 HD VET 9215 HD Working length 1000mm 1500mm 3300mm...

Page 17: ...deo endoscope into water to observe whether there are bubbles continuously bubbling up from the endoscope or not Normally 3 or less bubbles per minute is acceptable The continuous bubbles indicates le...

Page 18: ...sor isOFF Figure 3 2 Connect the power cord to power supply socket of endoscope imaging processor and to the wall mains outlet Turn on the power supply switch of the processor and assure the brightnes...

Page 19: ...e connector of the endoscope indicates it works normally Turn on the imaging processor and the monitor check whether the imaging color on the monitor is normal or not 3 3 1 Inspection of the insertion...

Page 20: ...the instrument channel valve has no any foreign objects or water drops Slowly insert the accessory through the biopsy valve and confirm that no any obstruction Do not force the biopsy forceps extend...

Page 21: ...bottle cap Turn ON the switch of the air pump built inside of the endoscope imaging processor Cover the air water feeding valve inapplicable to VET 3512 and confirm that air is fed to the distal end...

Page 22: ...nstrument channel confirm that the instrument channel is unobstructed and the biopsy forceps can smoothly extend from the distal end CAUTION Keep the suction pressure under 30kpa Excessive suction pre...

Page 23: ...r for automatic real coloradjustment Turn ON the LED lamp press the brightness adjustment buttons and adjust it to the appropriate observation intensity Feed air and water if necessary adjust the angu...

Page 24: ...028HD VET 8015 HD VET 9215 HD VET 3512 Figure 4 3 CAUTION AOHUA do not recommend the use of endoscopic instruments not approved by AOHUA Do not forcibly insert the instruments into the instrument chan...

Page 25: ...shut OFF and then disconnect the videoscope cable before disconnecting the endoscope from the imaging processor 4 5 Cleaning and disinfection of the endoscope After each examination disconnect the en...

Page 26: ...ine solution Glutaraldehyde solution 4 5 2 Preprocessing Before removing the endoscope from the endoscope imaging processor use a clean lint free cloth dampened with detergent to remove all visible di...

Page 27: ...distal end from the detergent and cover the suction valve to aspirate air for 10s Switch OFF the pump and disconnect the endoscope from the endoscope imaging processor remove suction valve and biopsy...

Page 28: ...lly the air water nozzle inapplicable to VET 3512 and the object lens ensure that all outer surfaces are completely cleaned According to the following steps brush the suction cylinder instrument chann...

Page 29: ...tep 2 and 3 at least 9 times to ensure that no debris is left To brush the suction cylinder route c 1 Insert the channel opening cleaning brush into the suction cylinder until half of the brush is ins...

Page 30: ...at the distal end of the endoscope is thoroughly immersed in the detergent 2 Use a 50ml syringe to respectively inject detergent into the air water channel inapplicable to VET 3512 and suction cylinde...

Page 31: ...urface of the endoscope air water valve inapplicable to VET 3512 suction valve and the biopsy valve with flowing water place them in a sterile container Use a 50ml syringe to respectively inject air i...

Page 32: ...infectant concentration recommended by the manufacturer Replace the gloves and use a 50ml syringe to respectively inject air into the air water cylinder inapplicable to VET 3512 and suction cylinder o...

Page 33: ...into the air water cylinder inapplicable to VET 3512 and suction cylinder through the injection tube at least 15 times 750ml Dry the outer surface of the endoscope the air water valve inapplicable to...

Page 34: ...imaging processor must be turned off before storage Ensure the power supply cord is disconnected The carrying case is not for storage do not use it to store the endoscope in case of contamination All...

Page 35: ...connector Insufficient angulation or angulation failure Check if the light fiber is broken Contact AOHUA Angulation lock failure Contact AOHUA Instrument fails to access the instrument channel Check...

Page 36: ...e the warranty card 5 2 Returning the endoscope for repair CAUTION AOHUA is not responsible for any injuries to the human or damages to the endoscope resulted from repair activities attempted by non A...

Page 37: ...e g sunlight emergent light of light source etc Damage may be caused to the light sensitive and precise device If the video endoscope does not function normally stop using it immediately and shut OFF...

Page 38: ...only Product Manufacture Date refer to the product nameplate Product service life 2 years Issue Date 2020 07 20 6 3 Manufacture Date CAUTION Daily check if the technical requirements of the endoscope...

Page 39: ...uct and contact Aohua maintenance center Be sure to send the warranty card of the product within one month after purchasing this product Warranty conditions Within half year after purchasing any quali...

Reviews: