For veterinary use only
-
16 -
Connect the monitor to the imaging processor using the BNC, S-VIDEO, and DVI cables, the ports and cables
can’t be altered.
−
Connect the video output terminal (AV) of the endoscope imaging processor with the video input
terminal (CVBS IN) of the monitor using the BNC Cable;
−
Connect the video output terminal (Y/C) of the endoscope imaging processor with the input
terminals (S-VIDEO IN) using the S-VIDEO Cable;
−
Connect the video output terminal (DVI) of the endoscope imaging processor with the input
terminals (DVI IN) using the DVI Cable.
CAUTION
⚫
For ancillary devices, such as recorders, printers, computer work stations etc., Consult
AOHUA to confirm equipment compatibility. AOHUA is not responsible for any consequences
caused by using of unapproved device.
⚫
When install the imaging processor, keep it away from equipment with high energy and high
power, or equipment emitting electromagnetic inference (such as CT equipment, X-ray
equipment, microwave therapy apparatus, high frequency hyperthermia instrument, MRI
equipment, radio communication device, etc.). Interruption of radio signals will appear on the
display screen if the inference occurs, resulting in degradation on image quality. It is
recommended to stay away from the inference source if the noise occurs.
⚫
Do not hit the terminals on the front and rear panels of the imaging processor with any hard
or sharp objects. Otherwise, damage can result.
⚫
Do not touch the electric connecting points in the output socket of the endoscope imaging
processor.
⚫
It is prohibited to use this product in the environment of flammable gas, otherwise a fire or
explosions may result.
⚫
It is strictly prohibited to frequently press the power switch of the imaging processor. Wait for
at least 8 seconds for next initiation.
⚫
Y/C terminal is used for connecting with the Y/C IN terminal of the compatible instrument.
When connecting with the devices mentioned above, the Y/C video connection cable of the
monitor can also be used, if more video connection cables are needed for connecting with
other devices, purchase more cables or use the video cables of other devices.
⚫
For the information of high frequency coagulation device compatible with this imaging
processor, refer to the instruction manual of the endoscope. Using the device not introduced
in the endoscope instruction manual may cause the monitor to produce high frequency
sound, and cause the endoscopic image to distort and disappear.