For veterinary use only
-
17 -
Confirm that the ventilation grills are not covered or blocked with dust or other materials, and that the imaging
processor can be turned ON.
Confirm that the ventilation grills are not covered or blocked with dust, and that the endoscope imaging
processor is turned ON.
Press the power switch of the endoscope imaging processor as shown in Figure 3.2. And the power
indicator lights up.
(Figure 3.2)
Confirm that air is ventilated by holding hand in front of the ventilation grills.
If the equipment fails to be turned ON, turn the endoscope imaging processor OFF.
For VET-OR1200HD and VET-OR1200R: Confirm that the power cord is firmly connected.
For VET-OR1200HD-Li: Confirm that the battery is normally working by checking the battery indicator as
described in the section 4.4.13.
Then, turn the endoscope imaging processor ON again. If the equipment still fails to be turned ON, contact
AOHUA.
Press the “Light” button as described in the section 4.4.2, “Examination lamp ignition” to ignite the
examination lamp. Confirm the examination light is emitted from the distal end of the endoscope.
3.2 Inspection of the power supply
⚫
If air is not ventilated through the ventilation grills, do not use the imaging processor and
contact AOHUA.
3.3 Inspection of the examination light
⚫
Do not look directly into the distal end of the endoscope when it is emitting light. Eye injury may
result.
1
2
3
4
1
5
6