04
Operation
For veterinary use only
-
20 -
The operator of the endoscope imaging processor must be a medical practitioner capable of safely performing
endoscopy after operation technique training. This manual does not explain or discuss clinical endoscopic
procedures. It only describes basic operation of the endoscope imaging processor. Before using the
endoscope, ensure to perform preparation and inspection on the endoscope imaging processor as described
in Chapter 3, “Installation and inspection”.
Turn the ancillary equipment ON.
Press the power switch to turn the endoscope imaging processor ON. The power indicator lights up.
Confirm that the endoscopic image shows on the monitor normally.
4.1 Precautions
⚫
If any irregularity is observed, do not use the endoscope imaging processor. Electric shock
may result.
⚫
Do not hit the terminals of the front and rear panel of the endoscope imaging processor with
sharp or hard objects. Otherwise, this instrument may be damaged.
⚫
While operating the endoscope and performing the endoscopy, be sure to observe the live
image. If operate the endoscope and perform the endoscopy while freezing or replaying the
images, patient injury may occur.
CAUTION
⚫
Before disconnecting the endoscope with the endoscope imaging processor, power supply
must be turned OFF.
⚫
It is prohibited to use the imaging processor in the environment of flammable gas, otherwise a
fire or explosions may result.
⚫
It is forbidden to frequently turn ON the power switch of the imaging processor. Wait for at least
8 seconds before turning the imaging processor ON after each shutdown.
4.2 Turning the endoscope imaging processor ON
NOTE
⚫
Also refer to instruction manuals of ancillary equipment for safe use.
1
2
3