15
12.2. RECORDING SETTINGS
Data is archived in a text file with a
csv
extension in the internal memory or the USB memory; a detailed
description of the storage format is given in chapter 14. Data is recorded until the memory is full (at least 300 days
of continuous operation with recording from 16 channels every 1 s) and then the recording is stopped and the
following message is displayed repeatedly “
Memory full. Recording rejected
” (chapter 15). The recording must then
be stopped (the
Recording type
must be set to the
Off
value, Table 12.2), the archive files must be copied for
further analysis, and space must be freed up in the memory for future recordings. The selection of the memory for
recording and the copying and deleting of existing
csv
files are accessible in the
Main Menu
->
Memory and file
options
; a detailed description of operations performed on files is given in chapter 12.3. The data recording
interval must be adjusted according to the needs of a specific application.
Table 12.2. Configuration parameters in the
Recording settings
menu
Parameter
Range of variability of the parameter and description
Default
settings
Data recording interval
1 s to 8 h (note 1 in the technical data, chapter 5)
(1)
1 minute
Recording type
(2)
(chapter 14)
Off
recording switched off
permanently
possibility to set those values using
the
[F]
button, chapter 10, item c
Disabled
Continuous
recording switched on
permanently
Limited by
date and time
recording is active within the range of the
Date
and the
Time
defined by the parameters
Start
and
End of time limit
Periodic
daily (hourly)
recording is active within the range of the
Time
defined by the
parameters
Start
and
End of time limit
Above the
permission
threshold
recording is active when the measured value of the channel
defined by the
Selection of permission channel
is larger than
the value of the
Permission threshold value
parameter
Below the
permission
threshold
recording is active when the measured value of the channel
defined by the
Selection of permission channel
is smaller than
the value of the
Permission threshold value
parameter
Start of time limit
Date
:
01.06.2008
÷
31.12.2099
,
Time
:
00:00:00
÷
23:59:59
, the parameter is
active when the
Recording type
=
Limited by date and time
or
Periodic daily
2013.06.01
00:00:00
End of time limit
Date
:
01.06.2008
÷
31.12.2099
,
Time
:
00:00:00
÷
23:59:59
, the parameter is
active when the
Recording type
=
Limited by date and time
or
Periodic daily
2013.06.01
00:00:00
Selection of permission
channel
Measurement channel 1
÷
Measurement channel 16
or
All channels
the parameter is active when
Recording type
=
Above
or
Below the
permission threshold
Measurement
channel 1
Permission threshold
value
-199.9
÷
1999.9°C
,
-9999
÷
19999
(3),
the parameter is active when
Recording
type
=
Above
or
Below the permission threshold
50.0 °C
Notes:
(1)
- the recording interval is counted from the moment the device is disconnected from the USB port
(2)
- the device does not record data in a file when it is connected to the computer's USB port
(3)
- applies to analog inputs (mA, V, mV, Ω) in the AR433/434 sensor
Icon
Type of menu item (parameter)
parameter that can be modified using the buttons and the touch screen, saved in the internal memory
informative item, not modified directly using the buttons and the touch screen
file or disk action (operation) (chapter 12.3)
Summary of Contents for AR407