background image

18

 

Assignment of  alarm 
outputs

 

combinations of

 Alarm outputs 1

 ÷ 

4

 or 

None

, selection of alarm outputs 

activated by the measurement of the current channel, chapter 12.6 

Alarm out. i 

(for i=1-4) 

Assignment of the 
measurement group 

channel 

Belongs to all

 groups or only to one group in the range of 

1

 

÷ 

8

channel grouping is used in the measurement presentation modes 

Group i 

(for 

i=1÷8)

 

Background color 

selection of the background color in the measurement presentation modes, 23 
colors are available 

color 

Notes: (1)

  - for radio sensors, the resolution of indications (point position) is taken remotely (by radio) 

 

      (2)

 – applies to analog inputs (mA, V, mV, Ω) in the AR433/434 sensor 

  

12.5. ADDING AND REMOVING OF MEASUREMENT SENSORS TO AND FROM A LIST 

As a default, the device is delivered with an empty list of measurement channels. For the purpose of presentation 
of data, new channels must be added to the list that are assigned to the radio sensor of the AR43x series and/or 
Apar devices connected to the RS485 line and/or the AR182/AR183 temperature probe. Selection of the serviced 
interface (radio, RS485, or probe) for each of the channels is set with parameter 

Assigned sensor type

 (in the 

Measurement channels 

configuration menu; a detailed description can be found in chapter 12.4).  

In a correctly configured system, all measurement channels of the same type (radio sensors or RS485 devices) 
should have their own unique identification numbers (ID or MODBUS-RTU address, set with the 

Address of 

assigned sensor ID

 parameter, chapter 12.4). 

When the 

Assigned sensor type

 = 

RS485 (MODBUS-RTU)

, it is also necessary to configure the 

Decimal dot 

position 

and the 

Register address to read for RS485

 (e.g. 16-bit measurement with a character, according to the 

map of MODBUS-RTU registers of the connected device that is available in its user’s instruction) and in the 

Communication settings

 (chapter 12.8), set the 

RS485 operation mode 

parameters to 

Master 

and the 

Baud rate 

for RS485

 to one that is compatible with the device.

 

If the 

Assigned sensor type

 = 

Radio (AR43x)

, then the measurement channel adding or deleting procedure 

follows the description below. 
Before the wireless sensors adding procedure is started, make sure that all devices of the configured system work 
on the same radio channel number that is different than that of the neighboring radio networks that are in the 
range (as a default - channel 6; the larger the number, the better; see chapter 12.8). 

1.  

There are two ways to add radio sensors to the list (indexing): 

a)

 

automatic (initiated with the 

ADD/REF/OFF

 button in the AR43x sensor) 

 

NOTE:  

 

Radio sensors 

must

 be indexed individually! 

The start 

of the procedure to obtain an index (

ID

) is indicated in the sensor by a single momentary 

illumination of the 

Status

 and 

RX/TX

 diodes, which takes place after the 

ADD/REF/OFF

 button is pressed.  

The end 

of the procedure takes place when the “

New radio sensor was registered

” or “

No available channels. 

Provide unused ones.

” message is displayed on the AR407 display. 

After the message is closed with the 

[SET]

 or 

[ESC]

 button, one can start indexing another sensor, unless the 

list is full (a maximum of 16 channels). In the event of a failure, the procedure must be repeated

. The 

maximum duration of the procedure is approximately 35 seconds

In the case of indexed 2-channel sensors (AR433/434/435/436/437), it is possible to delete the second channel 
from the list of the AR407 recorder (parameter 

Address of assigned sensor (ID)

 = 

None

=16 in the 

Measurement channel 

configuration menu, chapter 12.4), e.g. in order to switch this channel off or make it 

available to another sensor. 

 

NOTE:  

!

 

At the same time, if the procedure is started on several sensors, they are assigned the same 

ID

 address 

(number) and, as a result, they are assigned to the same measurement channels in AR407 (the data will be 

!

Summary of Contents for AR407

Page 1: ...USER INSTRUCTION AR407 16 CHANNEL RADIO AND WIRED DATA RECORDER Version 1 0 0 2017 04 10 93 87 10310 Tel 02 932 2326 Fax 02 903 0080 Ext 6867 Email info omi co th Website www omi co th...

Page 2: ...10 FUNCTIONS OF HARDWARE AND SCREEN BUTTONS 9 11 DESCRIPTION OF DATA SHOWN ON THE LCD DISPLAY 11 11 1 UPPER AND LOWER STATUS BAR 11 11 2 VIEW OF A MEASUREMENT GROUP TEXT ONLY 12 11 3 VIEW OF A MEASUR...

Page 3: ...e device d ensure proper operating conditions compliant with the technical specification of the device chapter 5 power supply voltage humidity temperature etc do not expose the device to direct and in...

Page 4: ...fox etc the site contains information about active measurement channels time status of the outputs recording etc with the possibility to show diagrams using the Google Chart API service diagrams requi...

Page 5: ...ltaneous recording of data from all active measurement channels an enclosure for panel installation leak tightness from the front side IP30 compliance with the R TTE 99 5 EC Directive possible update...

Page 6: ...ual COM serial port MODBUS RTU protocol master mode host support of USB memory pendrive up to 4 GB writing speed approx 135kB s depending on the memory RS485 galvanic separation MODBUS RTU protocol Ma...

Page 7: ...mbering and description of the connections on the back panel and method of connecting probes and electrical signals Terminal clamps connections Description ANTENNA SMA socket to connect an antenna for...

Page 8: ...OMx serial port x port number 1 2 etc The serial port uses the MODBUS RTU protocol In the internal memory two configuration text files are shown AR407 cfg and AR407 txt chapter 12 Communication with t...

Page 9: ...ord protection in the Access and other settings menu is switched on then the password must be entered chapter 12 9 closes the message window appearing on the screen chapter 15 UP or DOWN screen button...

Page 10: ...epending on the value of the F button function parameter in the Access and other settings menu chapter 12 9 Status signa ling icon Device status device status screen factory setting function available...

Page 11: ...g of the individual graphic elements is described below a lower status bar Fig 11 1 2 Lower status bar Object Description and marking in the contents of the instructions 6 no USB connection with the c...

Page 12: ...VIEW OF A MEASUREMENT GROUP TEXT ONLY Fig 11 2 Appearance of a measurement group screen in the TEXT ONLY mode presentation max 16 channels is possible 11 3 VIEW OF A MEASUREMENT GROUP TEXT AND A BAR...

Page 13: ...er configuration can be performed using one of the following three methods do not use them at the same time 1 From the film keypad and a touch screen located on the front panel of the device from the...

Page 14: ...e port and the MODBUS Address of the device in the case of the RS485 make sure that the serial port drivers have been installed correctly chapter 8 disconnect for a few seconds and then reconnect the...

Page 15: ...dic daily hourly recording is active within the range of the Time defined by the parameters Start and End of time limit Above the permission threshold recording is active when the measured value of th...

Page 16: ...rnal only archive files are created only in the internal memory Internal only Auto select archive files are created in the USB memory when detected or the internal memory when there is no USB memory A...

Page 17: ...e selected channel 1 16 Parameter Range of variability of the parameter and description Default settings Name unit and group edition of the name of the channel and the group max length 16 characters a...

Page 18: ...in its user s instruction and in the Communication settings chapter 12 8 set the RS485 operation mode parameters to Master and the Baud rate for RS485 to one that is compatible with the device If the...

Page 19: ...n be done in two ways a with the power supply of the AR407 recorder switched off or with the recorder outside of the radio range press the ADD REF OFF button in the sensor for about 10 sec the sensor...

Page 20: ...r 9999 19999 units 2 50 0 C Alarm hysteresis 1 0 500 0 C or 0 5000 units 2 0 0 C Output signal value duty cycle 1 100 1 step output control value for the activated alarm filling coefficient with the O...

Page 21: ...above in chapter 12 6 in Table 12 6 2 in item Email settings and status In order to eventually start the configured service the Email alarm notifications parameter must additionally be set to the Enab...

Page 22: ...IP number and router modem configuration To facilitate access to the networks with varying public IP address one can start the built in DDNS server service described in chapter 12 8 1 The selection o...

Page 23: ...network 192 168 0 10 DNS server 0 0 0 0 255 255 255 255 domain name server address DNS 192 168 0 10 Dynamic DNS server client DDNS Disabled A DDNS service that enables access over the Internet to a re...

Page 24: ...manent nature which indicates inadequate Internet connection improper configuration of the connection or service codes 2 to 12 e g DDNS 5 means invalid username or password DDNS 8 means invalid host n...

Page 25: ...ffices for at least 5 years of continuous operation Table 12 10 Configuration parameters in the Time and date menu Parameter Range of variability of the parameter Date yyyy mm dd 2008 06 01 2099 12 31...

Page 26: ...fg file off line NEW USER parameter configuration from the keypad and touch screen user level NEW CH_TEXT name configuration by way of modification of the AR407 txt file creation of a new csv file 4 I...

Page 27: ...s furniture etc b change the radio channel in all devices in the system in each separately to a different one c check the battery level in the AR43x sensor and if necessary replace the battery with a...

Page 28: ...line Fig 17 at the start of the line 2 x 820 to the ground and 5 V of the MASTER and 150 between lines at the end of the line 150 between lines termination and polarizing resistors when the MASTER is...

Page 29: ...0x84 0x02 0x0001 0x5130 19 MODBUS TCP SERIAL TRANSMISSION PROTOCOL The MODBUS TCP protocol is accessible to the Ethernet RJ45 interface and uses the TCP IP transport layer The parameters used by the s...

Page 30: ...s HB and hours LB R W 0x22 34 0x0000 0x3B3B minutes HB and seconds LB R W 0x23 35 0x0101 0x630C years HB and months LB Parameter End of the time limit chapter 12 2 R W 0x24 36 0x0100 0x1F17 days HB an...

Page 31: ...7 R W 0xFB 251 60 14400 Parameter Radio scanning interval chapter 12 8 number of seconds s R W 0xFC 252 0 6 Parameter Radio channel number chapter 12 8 R W 0xFD 253 0 1 Parameter USB operation mode ch...

Page 32: ...32 21 USER S NOTES...

Reviews: