12.7. DISPLAY SETTINGS
Table 12.7. Configuration parameters in the
Display settings
menu
Parameter
Range of variability of the parameter and description
Default
settings
Screen backlight time
0
÷
60 min.
, increase by 1 min., for the
0
value the illumination is on all the time,
the time is counted from the last use of the keypad or the touch screen
0 min.
(continuous)
Brightness of the screen 5
÷
100 %
, change by 5%
100%
Channel and group auto
switch time
0
÷
60 s
, change every 2 s, for the
0
value, the automatic change is off, the time
of automatic change of the channel (diagram) and the group in the
measurement presentation modes
0 sec.
Chart time range
100 s
300 s
15 min. 30 min. 60 min. 150 min. 5 h 10 h
25 h
100 s
50 h
5 days
10 days
Language
Polish
,
English
, the language of the menu (also covers the version of the
WWW server page)
Polish
12.8. RADIO, RS485, USB, AND ETHERNET COMMUNICATION SETTINGS
The available interfaces enable communication with Apar measurement devices and with other devices, e.g. a
computer, and, consequently, enable the readout of the measured values and configuration of the parameters and
the names, as well as access to the archive files (with the exception of the RS485). Moreover, the Ethernet enables
displaying information on the operating status and on measurements of the recorder in any web browser (Opera,
IE, Firefox, etc.) via the local network or the Internet and sending e-mail alarm notifications (chapter 12.6.1). The
WWW server uses the HTTP protocol on the standard port
80
. The Internet connection requires a known public IP
number and router (modem) configuration. To facilitate access to the networks with varying public IP address, one
can start the built-in DDNS server service (described in chapter 12.8.1).
The selection of the TCP and UDP port
number used by the recorder and the forwarding of this port in the router and other network parameters
configurations should be performed by a qualified person (the network administrator)
. Moreover, one must
make sure that the firewall does not block the ports and applications being used (ARSOFT-CFG and ARSOFT-LOG,
and the MODBUS-TCP protocol). Additional information about the RS485 interface and the MODBUS-RTU/TCP
protocols is available in chapters 17
÷
20.
Table 12.8. Configuration parameters in the
Communication settings
menu
Parameter
Range of variability of the parameter and description
Default
settings
Radio scanning
interval
1 min.
÷
4 hours, from 2 min. in the case of radio channels 4
÷
6, change every 1
min.
5 min.
Radio channel
number
0
÷
6
the selected channel must be set separately in each device
(radio sensor) operating in the system
6
USB operation mode
Available for
a computer
(device)
(2)
in order to establish communication with the computer,
drivers must be installed (chapter 8); the USB connection is
indicated in the lower status bar (chapter 11.1, item b)
Available for
a computer
USB memory
support (host)
the presence of the USB memory is indicated in the lower
status bar (chapter 11.1),
do not connect to the
computer's USB port
RS485 operation
mode (MODBUS-RTU)
Master
to read measurements from APAR devices (chapter 12.5)
Slave
Slave
for communication with a master device, e.g. a computer
Baud rate for RS485
2,400 bit/s
4,800 bit/s
9,600 bit/s
19,200 bit/s
19,200 bit/s
38,400 bit/s
57,600 bit/s
115,200 bit/s