26
Appendix I: Parameter setting and factory default setting
Modes
Prompts
Parameter setting items
Abbreviation
Description
Restore
to factory
default
setting
P1.0
pH
P1.1
Select pH buffer solution
USA
-
NIST
-
CUS
USA
P1.2
Select resolution
0.01
-
0.1
0.01
P1.3
Set reading stability
criteria
Normal
-
High
-
Low
Normal
P1.4
Set due calibration
No
-
H00
-
D00
No
P1.5
Check the date of the last
calibration
/
-
-
P1.6
Restore factory default
setting
No
-
Yes
No
P2.0
Conductivity
P2.1
Select electrode constant
1.0
-
10.0
-
0.1
1.0
P2.2
Select conductivity
standard solution
Standard
—
CUS
Standard
P2.3
Select reference
temperature
(15~30)
℃
25
℃
P2.4
Adjust temperature
compensation coefficient
0.00~9.99
2.00
P2.5
Adjust TDS factor
0.40~1.00
0.71
P2.6
Set due calibration
No
-
H00
-
D00
No
P2.7
Check the date of the last
calibration
/
-
-
P2.8
Restore factory default
setting
No
-
Yes
No
P3.0 Basic
parameters
P3.1
Select temperature unit
/
°C - °F
°C
P3.2
Select back light timing
1
-
2
-
3
-
On
1
P3.3
Adjust storage timing
/
-
0:00
P3.4
Adjust date
/
-
-
P3.5
Adjust time
/
-
-
P3.6
Clear stored data
No
-
Yes
No
P3.7
Automatic Power-off setup
10
-
20
-
30
-
On
20
P3.8
Auto. Hold
/
Off
-
On
Off