4.4.3. pH temperature principle 

Please note that the closer the temperature of the sample solution to the calibration solution, the more 

accurate the readings.   

 

4.4.4. Factory default setting 

For factory default setting, please refer to parameter setting P1.4 (Item 7.3). All calibration data is deleted 

and the meter restores to the theory value (zero electric potential of pH is 7.00, the slope is 100%). Some 

functions restore to the original value (refer to appendix -1). When calibration or measurement fails, please 

restore the meter to factory default setting and then perform re-calibration or measurement. Please note 

once set the factory default, all the data deleted will be irretrievable. 
 

4.5  pH Electrode Maintenance 

4.5.1. Daily maintenance 

The soaking solution contained in the supplied protective bottle is used to maintain activation in the glass 

bulb and junction. Loosen the capsule, remove the electrode and rinse in distilled or deionized water before 

taking a measurement. Insert the electrode and tighten the capsule after measurements to prevent the 

solution from leaking. If the soaking solution is turbid or moldy, replace the solution. The soaking/storage 

solution is 3M KCL (SKU: AI1107). Using other brand’s storage solution might cause potential damage to 

the electrode.   

 

The electrode should not be stored in distilled or deionized water, protein solution or acid fluoride solution. 

In addition, do not soak the electrode in organic silicon lipids. For best accuracy, always keep the meter 

clean and dry, especially the meter’s electrode socket. Clean with medical cotton and alcohol if necessary. 
 
4.5.2. Calibration buffer solution 

For calibration accuracy, the pH of the standard buffer solution must be reliable. The buffer solution should 

be refreshed often, especially after heavy use. 
 
4.5.3. Protect glass bulb   
The sensitive glass bulb at the front of the combination electrode should not come in contact with hard 

surfaces.  Scratches  or  cracks  on  the  electrode  will  cause  inaccurate  readings.  Before  and  after  each 

measurement, the electrode should be washed with distilled or deionized water and dried. Do not clean the 

glass bulb with a tissue for it will affect the stability of the electrode potential and increase the response 

time. The electrode should be thoroughly cleaned if a sample sticks to the electrode. Use a solvent if the 

solution does not appear clean after washing. 
 
4.5.4. Renew glass bulb 
Electrodes that have been used over a long period of time will become aged. Submerge the electrode in 

0.1mol/L  hydrofluoric  acid  for  24  hours,  then  wash  the  electrode  in  distilled  or  deionized  water,  then 

submerge it in soaking solution for 24 hours. The method to prepare 0.1mol/L hydrofluoric acid dilute 9mL 

hydrochloric acid in distilled or deionized water to 1000mL. For serious passivation, submerge the bulb in 

Summary of Contents for PH850

Page 1: ...Version2 0 PH850 Portable pH Meter User Manual...

Page 2: ...3 6 Automatic Power off 5 3 7 Manual Temperature Compensation MTC 5 4 pH Measurement 5 4 1 pH Electrode Information 5 4 2 pH Calibration Consideration 5 4 3 pH Meter Calibration with an example of 3 p...

Page 3: ...ssor based portable meter features automatic calibration automatic temperature compensation function setup self diagnostics automatic power off and low voltage display The meter s digital filter impro...

Page 4: ...manual or automatic mV Range 1 000 mV Resolution 1 mV Accuracy 0 2 F S 1 digit Temperature Range 0 100 C Resolution 0 1 C Accuracy 0 5 C 1 digit 2 2 Other Specifications Power AAA batteries 3 1 5V 3 I...

Page 5: ...al temperature compensation 10 Calibration guide icon 11 Stability icon of readings 12 Low battery icon when this icon appears please replace the battery 13 Automatic reading hold icon 3 2 Keypad Func...

Page 6: ...this key to enter the calibration mode In the calibration mode or the parameter setup and auto lock up HOLD mode press this key to return to the measurement mode Short press In the measurement mode pr...

Page 7: ...C 4 pH Measurement 4 1 pH Electrode Information The meter connects to 201T F plastic three in one combination pH electrode with built in temperature sensor with automatic temperature compensation func...

Page 8: ...ation Accuracy 0 1pH Two point calibration Range 7 00pH 0 Range 7 00pH Three point calibration 0 Large Range 4 2 3 Calibration frequency Calibration frequency depends on the sample the electrode perfo...

Page 9: ...on LCD press key to calibrate the meter LCD will display electrode slope of acidity range then 2nd point calibration is finished the meter enters measurement mode of two point calibration Calibration...

Page 10: ...r connects the electrode well 3 Check whether the electrode is damaged Pressing key too soon Press key after icon appears and stays During calibration the measuring value is not stable for 3min 1 Chec...

Page 11: ...d in distilled or deionized water protein solution or acid fluoride solution In addition do not soak the electrode in organic silicon lipids For best accuracy always keep the meter clean and dry espec...

Page 12: ...on stir the solution briefly and allow it to stay in the solution until icon stays and then record the reading which is ORP value ORP is in short for Oxidation Reduction Potential The unit is mV 5 2 N...

Page 13: ...r submenu P1 1 of pH parameter setting then press key to switch among submenu P1 1 P1 2 P1 3 P1 4 see Diagram 9 7 2 3 In P3 0 mode press key to enter submenu P3 1 of basic parameter setting then press...

Page 14: ...ia Normal High Low 1 Press key nor flashes Press key to select nor HI Lo press to confirm Nor Normal Hi High Lo Low 2 After confirming parameter press key to enter P1 4 mode or press key to return to...

Page 15: ...Automatic lock up setup Off On 1 Press key OFF flashes press key to select OFF On press to confirm Off not set On set the reading is automatically locked when stabilizes 10 seconds 2 When Parameter i...

Page 16: ...21 High Temperature solution LabSen 861 Low temperature solution LabSen 881 Meat LabSen 763 Micro volume solution LabSen 241 6 LabSen 241 3 Purified Water Low ion concentration samples LabSen 803 LabS...

Page 17: ...sibility of APERA INSTRUMENTS LLC for a period of THREE YEARS for the instrument and SIX MONTHS for the probe from the delivery This limited warranty does NOT cover any issues due to Accidental damage...

Page 18: ...iteria Normal High Low Normal P1 4 Restore to factory default setting No Yes No P2 0 Cond P2 1 Select electrode constant 1 0 10 0 0 1 1 0 P2 2 Select reference temperature 15 30 1 P2 3 Adjust temperat...

Page 19: ...t up reading stability criteria P1 4 Factory default setting Factory default setting P2 0 Conductivity P2 1 Cell Constant Cell P2 2 Reference temperature Reference temperature P2 3 Temperature compens...

Page 20: ...18...

Reviews: