1413 μS/cm. Use the equipped 2301T-S conductivity electrode (K = 1.0). Please choose the
calibration solution according to the Table (6-1).
2)
The conductivity electrode has been calibrated before meter leaving factory. In general, user can
directly use it.
3)
We recommend calibrating the electrode once every month. It is necessary to perform calibration
when replacing a new conductivity electrode.
Table (6-1)
Measurement range
0.05 to 20μS/cm
0.5μS/cm to 200mS/cm
Electrode constant
K=0.1 (flow test)
K=1.0
Calibration solution
84μS/cm
84μS/cm
1413μS/cm
12.88 mS/cm
111.9 mS/cm
Calibration icon
Note: When you choose to use DJS-0.1-S pure water conductivity electrode (sold separately, K=0.1, no temperature sensor
built in), you can use Manual Temperature Compensation mode: long press , temperature value starts flickering,
press or to change the value of temperature and then press to confirm.)
8.6
Parameter Setting:
Conductivity measurement parameter setting Table
Table (6-3)
Prompt Mark
Parameter Setting Items
Code
Parameter
P1
Standard solution series selection
USA (84μS/cm, 1413μS/cm, 12.88
mS/cm, 111.9 mS/cm)
CH (146.6μS/cm, 1408μS/cm,
12.85mS/cm, 111.3mS/cm)
P2
Electrode constant selection
0.1, 1, 10
P3
Reference temperature selection
25˚C 20˚C 18˚C
P4
Temperature compensation
coefficient setting
0.00 to 9.99%
P5
Electrode constant setting
P6
Temperature unit setting
˚C / ˚F
P7
Backlight display time setting
0-1-3-6min
P8
Auto power off time setting
0-10-20min
P9
Restore to producer setting
OFF-On