- 27 -
9.9
Parameter Setting
9.9.1 DO parameter settings (Chart 4-1)
Prompt Mark
Parameter Setting Items
Code
Parameter
P1
Resolution selection
0.01/0.1(mg/L and ppm)
0.1/1(%)
P2
Salinity calibration
P3
Barometric pressure setting
P4
Temperature unit setting
˚F/˚C
P5
Backlight display time setting
0-1-3-6 min
P6
Auto power off time setting
0-10-20 min
P7
Restore to producer setting
OFF-On (shut-set)
9.9.2 Resolution selection (P1)
(a) Press
to enter P1 mode, see picture (4-2);
(b) Press
to choose resolution: 0.01→0.1;
(c) Press
to enter the next parameter setting or press
to
return to the measurement mode.
9.9.3 Salinity calibration setting (P2)
(a) In P1, short press
to enter P2, see picture (4-3).
(b) Immerse the DO electrode into 12.88mS/cm calibration solution
(sold separately), solution level should exceed
salinity electrode; let
it stand still after stirring; then press
when the value is stable
and smiley face icon appears.
12.9
will show up on the screen. The
calibration will be finished after several seconds, and the meter will display a stable salinity value.
(c) Short press
to enter the next parameter setting or short press
to return to
measurement mode.