20
21
English
Electrical.fast.
transient/burst.
IEC.61000-4-4
±2.kV.for.power.
supply.lines
±1.kV.for.input/output.
lines
±2.kV.for.power.
supply.lines
±1.kV.for.input/output.
lines
Mains.power.quality.should.be.that.
of.a.typical.commercial.or.hospital.
environment.
Surge.
IEC.61000-4-5
±1.kV.differential.
mode
±2.kV.common.mode
±1.kV.differential.
mode
±2.kV.common.mode
Mains.power.quality.should.be.that.
of.a.typical.commercial.or.hospital.
environment.
Voltage.dips,.short.
interruptions.and.
voltage.variations.
on.power.supply.
input.lines
IEC.61000-4-11
<.5.%.UT.
(>95.%.dip.in.UT).for.
0.5.cycle
40.%.UT
(60.%.dip.in.UT).for.5.
cycles
70.%.UT
(30.%.dip.in.UT).for.
25.cycles
<5.%.UT
(>95.%.dip.in.UT).for.
5.sec
<5.%.UT.
(>95.%.dip.in.UT).for.
0.5.cycle
40.%.UT
(60.%.dip.in.UT).for.5.
cycles
70.%.UT
(30.%.dip.in.UT).for.25.
cycles
<5.%.UT
(>95.%.dip.in.UT).for.
5.sec
Mains.power.quality.should.be.that.
of.a.typical.commercial.or.hospital.
environment.
If.the.user.of.the.EQUIPMENT.or.
SYSTEM.requires.continued.operation.
during.power.mains.interruptions,.it.is.
recommended.that.the.EQUIPMENT.
or.SYSTEM.be.powered.from.an.
uninterruptible.power.supply.or.a.
battery.
Power.frequency.
(50/60.Hz).
magnetic.��
IEC.61000-4-8
3.A/m
3.A/m
Power.frequency.magnetic.���.should.
be.at.levels.characteristic.of.a.typical.
location.in.a.typical.commercial.or.
hospital.environment.
DIGI STIM_3 languages .indd 21
2013/2/23 AM 09:12:15
Summary of Contents for DIGI-STIM
Page 1: ...Model GM320T SD TENS DIGI STIM_3 languages indd 1 2013 2 23 AM 09 12 12 ...
Page 27: ...25 English DIGI STIM_3 languages indd 25 2013 2 23 AM 09 12 16 ...
Page 53: ...Español 51 DIGI STIM_3 languages indd 51 2013 2 23 AM 09 12 18 ...
Page 77: ...Português 75 DIGI STIM_3 languages indd 75 2013 2 23 AM 09 12 20 ...
Page 80: ...Português 78 DIGI STIM_3 languages indd 78 2013 2 23 AM 09 12 20 ...