the control solution at the edge of the yellow reaction zone.
Step 5:
The display screen will show timing bars"-----". The bars will flash and gradually
diminish for countdown.
Step 6:
After the timing bars disappear, the screen will show test result. Compare the
reading on the screen to the Level 1 range printed on the test strip package. The
test result falling within the range means the meter and the test strip are working
together properly. Remove the test strip.
Step 7:
To perform test of Level 2 control solution (solution with blue bottle cap), repeat
steps 1-6.
IMPORTANT:
If the test result is not within the expected range, repeat the control solution testing.
If the results continue falling outside the expected range, call our authorized dealer.
16
USER'S GUIDE
Chapter 2.
Before Testing