32
USER'S GUIDE
Chapter 8.
Message Interpretations & What to Do
Message
Meaning
What To Do
"
"
The temperature of your
The meter needs to be moved to an
meter, the surrounding air,
area that is within the system's
or the test strip is outside of operation range of 18°C to 38°C
system's operation range.
(64°F to 100°F), and wait for 10~15
minutes to let the meter adjust to the
new temperature.
Code number and
Not enongh blood sample
First check if the yellow reaction zone
flashing "
"
is applied to the yellow
is filled with blood sample, if it is not,
displaying after
reaction zone.
discard the test strip from the meter
blood sample
and retest with a new test strip.
is applied.
LO
The blood glucose result
Check using right code card and
is lower than 30mg/dL
new test strip, and the blood volume
(1.7mmol/L)
is suitable for applying, then repeat
the test to confirm the test result. If it
reads LO again, contact your
healthcare professional.
Code number and
flashing "
"
displaying after blood
sample is applied.