23
d.
Click on "Close" when adjustment is done. Click on "Yes" when the message window pops
up. The measured height will then be assigned to selected area.
Figure 27: Assign measured height to selected area.
Figure 28: Scan height shown in the "Scan Area Properties" pane.
e.
User can type in the height number to other scan area heights if these scan areas are at the
same height.
4.5.
Calibrate Probe XY Offset
Although the whole system has been calibrated carefully, it is still possible to have misalignment
between the probe axis and the robot shaft's axis, which will show mismatch between the
assigned scan points and real probe landing positions. The XY offset is then introduced to
compensate for this position misalignment.