72
AT4610/AT4710 User’s Guide
TC-K
Table 12-3 Specifications for thermocouple TC-K
Range(f.s.)
MAX Resolution
Measurement Range
Accuracy
100 °C
0.01 °C
-100.00 ~ 0 °C
±0.8 °C
0 ~ 100 °C
±0.6 °C
500 °C
0.05 °C
-270.00 ~ -100.00 °C
±1.5°C
-100 ~ 0 °C
±0.8 °C
0 ~ 500 °C
±0.6 °C
2000 °C
0.1 °C
-270.0 ~-100 °C
±1.5 °C
-100 ~ 1300 °C
±0.8°C
TC-J
Table 12-4 Specifications for thermocouple TC-J
Range(f.s.)
MAX Resolution
Measurement Range
Accuracy
100 °C
0.01 °C
-100.00 ~ 0 °C
±0.8 °C
0 ~ 100 °C
±0.6 °C
500 °C
0.05 °C
-210.00 ~ -100.00 °C
±1.0 °C
-100 ~ 0 °C
±0.8 °C
0 ~ 500 °C
±0.6 °C
2000 °C
0.1 °C
-210.0 ~-100 °C
±1.0 °C
-100 ~ 1200 °C
±0.8 °C
TC-E
Table 12-5 Specifications for thermocouple TC-E
Range(f.s.)
MAX Resolution
Measurement Range
Accuracy
100 °C
0.01 °C
-100.00 ~ 0 °C
±0.8 °C
0 ~ 100 °C
±0.6 °C
500 °C
0.05 °C
-200.00 ~ -100.00 °C
±1.0 °C
-100 ~ 0 °C
±0.8 °C
0 ~ 500 °C
±0.6 °C
2000 °C
0.1 °C
-200.0 ~-100 °C
±1.0 °C
-100 ~ 1000 °C
±0.8 °C
TC-T
Table 12-6 Specifications for thermocouple TC-T
Range(f.s.)
MAX Resolution
Measurement Range
Accuracy
100 °C
0.01 °C
-100.00 ~ 0 °C
±0.8 °C
0 ~ 100 °C
±0.6 °C
500 °C
0.05 °C
-270.00 ~ -100.00 °C
±1.5 °C
-100 ~ 0 °C
±0.8 °C
0 ~ 400 °C
±0.6 °C
2000 °C
0.1 °C
-270.00 ~ -100.00 °C
±1.5 °C
-100 ~ 0 °C
±0.8 °C
0 ~ 400 °C
±0.6 °C
TC-N
Table 12-7 Specifications for thermocouple TC-N
Range(f.s.)
MAX Resolution
Measurement Range
Accuracy
100 °C
0.01 °C
-100.00 ~ 0 °C
±1.2 °C
0 ~ 100 °C
±1.0 °C
500 °C
0.05 °C
-200.00 ~ -100.00 °C
±2.2 °C
-100 ~ 0 °C
±1.2 °C
0 ~ 500 °C
±1.0 °C
2000 °C
0.1 °C
-200.00 ~ -100.00 °C
±2.2 °C
-100 ~ 0 °C
±1.2 °C
0 ~ 1300 °C
±1.0 °C