Cleaning Procedure:
1.
Shut down to RO system.
2.
Disconnect permeate line and divert to drain before
any cleaning cartridge is installed.
3.
Remove the filter cartridge from the pre-filter housing.
4.
Replace the filter cartridge with the cleaning cartridge
and assemble into the filter housing.
5.
Turn system ON. After 30-40 seconds*, shut down
the system.
6.
Let the membrane(s) soak in the cleaning solution
overnight.
7.
Remove the empty cleaning cartridge and replace it
with the original filter.
8.
Restart the system. Direct the permeate to drain for 5
minutes.
9.
Go back to normal operations.
*
Instead of time, you may use one of the following criteria:
a.
Run the system until the pH of the concentrate is
almost the same as the cleaning solution (pH=3)
b.
Permeate rate for the system drops to a very low
value.
C-C2510-A11
Acidic Cleaning Cartridges for TF Type Membranes
User Instructions
Caution: Handle cartridges carefully.