94
reset back to the set screen.
5.2.2 DC Output
When the AUTO RUN parameter in the System Parameters menu is set to
PROGRAM Mode the Set screen will be displayed as follows:
To initialize the test press the Test/Reset key and the LED for the key will
illuminate. The text “Set” on the set screen will turn to Dwell, in addition your soft
keys will change to include Meter and Keypad. The output will run continuously
until the Test/Reset key is pressed again, or there is a failure condition.
When the RESET key is pressed or a Hi-Limit failure occurs, the instrument DC
output voltage has a fixed 1 second ramp down discharge time. The Test/Reset
LED flashes during this ramp down period to indicate the output is still active.
If you press the Meter soft key a shaded black box (■) will highlight the meter
parameters P: or A: and the display will read the output on the right side of the
display. Every time the meter key is pressed it will toggle between the two meter
parameters.
If you press the Keypad soft key the display will show the text Voltage = above
the voltage meter on the left hand side of the display. A shaded black box (■) will
be flashing waiting for a voltage value to be entered from the numeric keypad.
Once the value has been typed into the instrument you must press the Enter soft
key to accept the value. The Esc soft key is available to exit out of this mode and
Summary of Contents for 00310XAC
Page 16: ...10 310XAC 320XAC...
Page 24: ...18 Accuracy L 0 1 Hz 0 0 500 Hz H 0 2 Hz 501 1000 Hz...
Page 26: ...20 A...
Page 36: ...30...
Page 101: ...95 move back to the test screen...
Page 108: ...102...
Page 127: ...121 CONNECT CONNECT 0 1 0 1...
Page 130: ...124 Register...
Page 138: ...132...