114
9.4.1 Basic Commands and Query Commands
The following commands are used to control actual output voltage and current
from the instrument. This command set also includes query commands. These
query commands will retrieve data from the instrument. The GPIB bus application
requires an IEEE-488 read command to be sent after the query command. These
commands include functions for retrieving test data, test results and metering
values.
Command
Description
AC Program
Mode
AC Manual
Mode
DC Program
Mode
DC Manual
Mode
Unit
TEST
TEST
Power On
RESET
RESET
Power Off
TD?
Testing meters
data
Memory, Step, Status,
Freq, Volt, Curr,
Power, APeak, PF, Q,
CF, VA, Timer
Memory, Status,
Freq, Volt, Curr,
Power, APeak, PF,
Q, CF,VA
Memory, Step, Status,
Volt, Curr, Power,
Timer
Memory, Status,
Volt, Curr, Power
RD XX?
Results meters
data
TDFREQ?
Testing
frequency meter
Hz
TDVOLT?
Testing voltage
meter
5.0~300.0
5.0~420.0
V
TDCURR?
Testing current
meter
See Note1
See Note1
See Note2
See Note2
A
TDAP?
Testing current
peak meter
See Note4
A
TDP?
Testing power
meter
See Note3
W
TDPF?
Testing pf meter
0.000~1.000
TDQ?
Testing Q meter
See Note3
TDCF?
Testing CF
meter
0.00~10.00
TDVA?
Testing VA
meter
See Note3
VAR
TDTIMER?
Testing timer
meter
0.0~999.9
0.0~999.9
s/m/h
METER X
SELECT
METER X
X=0-8,0=F,1=A,2=P,
3=Q,4=CF,5=AP,6=V
A,7=PF,8=CYCLE
X=0-
7,0=F,1=A,2=P,
3=Q,4=CF,5=AP,6
=VA,7=PF
X=1,2,8
1=A,2=P,8=CYCLE
X=1,2
1=A,2=P
METER?
SELECT
METER?
SDTRG
TRANSIENT
TRIGGER
Trigger one time
TRANSIENT
STEPCYCLE?
STEP CYCLE?
MEMORYCYCLE?
MEMORY
CYCLE?
LOOPCYCLE?
LOOP CYCLE?
TEST
Turns on the output voltage at the selected step loaded into memory.
Summary of Contents for 00310XAC
Page 16: ...10 310XAC 320XAC...
Page 24: ...18 Accuracy L 0 1 Hz 0 0 500 Hz H 0 2 Hz 501 1000 Hz...
Page 26: ...20 A...
Page 36: ...30...
Page 101: ...95 move back to the test screen...
Page 108: ...102...
Page 127: ...121 CONNECT CONNECT 0 1 0 1...
Page 130: ...124 Register...
Page 138: ...132...