149
PRODUCT#.
The Autostart feature will not work with the SERIAL# setting.
The Autostart feature will enable the instrument’s output once the product
number barcode is scanned a second time when in the PRODUCT# and
SER/PROD modes. Do not touch the DUT at any time when using this feature in
order to avoid potential shock or serious injury.
Use the Change soft key to select the Autostart setting. Press the Enter soft key
to accept the new setting or the Exit soft key to cancel and return to the original
setting.
11.4.13
Ethernet Card Settings Commands and Companion Queries
Command
Name
Value
SIM {1|0}
SIM?
Set IP Mode
1=Manual, 0=Auto
(DHCP/BOOTP)
SIA <value>
SIA?
Set IP Address
Dotted decimal form. Ex.
192.168.1.50
SGA <value>
SGA?
Set Gateway IP
Address
Dotted decimal form
SSM <value>
SSM?
Set Subnet Mask
Dotted decimal form
SDN <value>
SDN?
Set Device Name
8 character max, must start with
a letter
MAC?
MAC Address Query
Example response:
00:20:4A:8B:B4:30
SBI {4|3|2|1|0}
SBI?
Set Barcode Input
0=Off, 1=Serial# and Product#,
2=Serial# Only, 3=Product#
Only, 4=Run File
SAS {1|0}
SAS?
Set Autostart
1=On, 0=Off
Communication Considerations
All of the above commands (excluding the query commands) will respond
with the 06 hex (6 decimal) Acknowledge (ACK) ASCII control code if the
transfer was recognized by the instrument.
If there was an error with the command string, the instrument will respond
with 15 hex (21 decimal), the Not Acknowledge (NAK) ASCII control code.
Summary of Contents for 00310XAC
Page 16: ...10 310XAC 320XAC...
Page 24: ...18 Accuracy L 0 1 Hz 0 0 500 Hz H 0 2 Hz 501 1000 Hz...
Page 26: ...20 A...
Page 36: ...30...
Page 101: ...95 move back to the test screen...
Page 108: ...102...
Page 127: ...121 CONNECT CONNECT 0 1 0 1...
Page 130: ...124 Register...
Page 138: ...132...