115
FHI?
FREQ HI?
40.0~1000
Hz
FLO XXXX
FREQ LO XXXX
XXXX=40.0~1000
Hz
FLO?
FREQ LO?
40.0~1000
Hz
SAG XXXX
START ANGLE
XXXX
XXXX=0-359
°
SAG?
START ANGLE?
0-359
°
EAG XXXX
END ANGLE
XXXX
XXXX=0-359
°
EAG?
END ANGLE?
0-359
°
RESULTS X
RESULTS X
X=0-2, 0=ALL,1=P/F,2=LAST
°
RESULTS?
RESULTS?
0-2
°
OF X
OC Fold X
X=0~1, 0=OFF,1=ON
OF?
OC Fold?
0~1
SD X
TRANSIENT X
X=0~1,0=OFF,1=ON
SD?
TRANSIENT?
0~1
LOCK X
LOCK X
X=0~1, 0=OFF,1=ON
LOCK?
LOCK?
0-1
MEMLOCK X
MEMLOCK X
X=0~1, 0=OFF,1=ON
MEMLOCK?
MEMLOCK?
0-1
VS X
VOLT SENSE X
X=0~1, 0=INT,1=EXT
VS?
VOLT SENSE?
0~1
FUNCTION X
FUNCTION X
Two units link X=2|3, Three units link X=1|2
Two units link X=2|3, Three units link X=2
FUNCTION?
FUNCTION?
1=3Ø4W, 2=Parallel, 3=1Ø3W
SSI X
SYNC SIGNAL X
X=0~3, 0=OFF,1=START,2=EVENT,3=FREQ
X=0~2, 0=OFF,1=START,2=EVENT
SSI?
SYNC SIGNAL?
0~3
0~2
9.4.6 IEEE 488.2 Common Commands
These commands are required by the IEEE-488.2 standard with the exception of
*PSC, *PSC?. Most of these commands are not available over the RS-232 bus
except for the *IDN? command which can be used to retrieve the instrument
identification information, and the four status reporting commands *ESR?, *ESE,
*ESE? and *STB?.
Command Name
Description
*IDN?
Identification Query
Company, Model Number, Serial
Number, Firmware Revision
*RST
Reset Command
Resets Unit
*TST?
Self-Test Query
00H=OK
01H=TEST EEPROM ERROR
*CLS
Clear Status Command
Clear Standard Event Status
Register Clear Service Request
Register
*OPC
Operation Complete Command
When TEST command ok setting
ESR BIT0 =1
*OPC?
Operation Complete Query
0 = Test in Process
Summary of Contents for 310XAC
Page 13: ...13 320XAC 340XAC...
Page 116: ...116 1 Test Complete OK WAI Wait for next command...