139
accept the new setting or the EXIT key to cancel and return to the original
setting.
11.4.12
Autostart
Highlight the Autostart parameter using the
∧
,
∨
soft keys. When the Autostart
parameter is highlighted, press the Edit soft key.
When Autostart is enabled, the test will execute as follows:
If the Barcode INPUT is set to PRODUCT#, scan the barcode once to input it into
the APT 300XAC. The APT 300XAC will then search for a test file name that
matches the product number barcode string. If the APT 300XAC finds a match, it
will load the file into RAM.
When the same product number barcode is scanned a second
time, the test will be executed automatically. If APT 300XAC does not find a file
name that matches the barcode string, the unit will beep – notifying the user that
it did not find a matching file name. The test file name is limited to 10 characters.
However, if the user names a test file with the maximum 10 characters, this
function will still load a test file if the first 10 characters of the product number
match the file name.
If the Barcode INPUT is set to SER/PROD, scan the serial number once to input
it into the APT 300XAC. Next, scan the product number. From this point, APT
300XAC will operate the same as when the Barcode INPUT setting is set to
PRODUCT#.
The Autostart feature will not work with the SERIAL# setting.
The Autostart feature will enable the instrument’
s output once the product
number barcode is scanned a second time when in the PRODUCT# and
SER/PROD modes. Do not touch the DUT at any time when using this feature in
order to avoid potential shock or serious injury.
Use the Change soft key to select the Autostart setting. Press the Enter soft key
to accept the new setting or the Exit soft key to cancel and return to the original
setting.
WARNING