38
function to work, the Lock must first be turned ON. Selecting the Mem Lock OFF
will allow the user to access all available memory locations but restricts access to
memory and step editing capabilities. Selecting the Mem Lock ON will allow the
user to only run the currently loaded memory.
4.2 System Parameters Description
The system parameters change the overall operation of the AC power source. If
the operator elects to edit the system parameters this will apply a universal
change to every memory and step location for the AC power source when in the
Test Parameters menu. The operator cannot independently change these
settings from one memory or step location to another.
1.
Auto Run
- places the AC power source into one of two modes
(PROGRAM/MANUAL). In the PROGRAM Mode the operator will have the
ability to program individual memories and steps with user selectable testing
parameters such as test time, high and low limits, etc.
In the MANUAL Mode the operator will have limited choices in selecting and
editing testing parameters. The key difference is that in MANUAL Mode the
operator cannot connect steps or have fixed testing times. The output is
either ON or OFF in the MANUAL Mode.
2.
Out Mode
- configures the instrument for AC or DC voltage output.
3.
Single Step (PROGRAM mode)
- controls how the instrument will
proceed from one test step to the next in a testing sequence when in the Auto
Run mode of PROGRAM.
When this parameter is selected on the source will sequence from one test
step to the next only when the Test/Reset key is pressed between each step.
When the Single Step parameter is on the source will pause after each step
has completed a test routine and passed based on the programmed testing
parameters. If a PASS occurs for the step the operator can proceed to the
next step in the sequence. If a FAIL occurs for the step the operator will not
be able to proceed in the test sequence. They will have to restart from the
beginning of the test sequence or step number one.
If the parameter is selected off the source will automatically sequence from
one step the next regardless if a pass or failure has occurred for a particular
step.
4.
Alarm
- controls the volume level of the alarm if a failure is detected. This
setting is from 1 – 9 with 9 being the loudest volume level.