61
halfway point (8 millisecond) of the sine wave or at the 180º point. Trans-Site
is only selectable when the source is configured for AC output.
31.
Trans-Time
- gives the operator the flexibility to program the overall time
duration of the surge or drop voltage. For example if the Trans-Site is 8
milliseconds; the output voltage is 120 volts; the surge voltage is 150 volts;
and the Trans-Time is 20 milliseconds when the sine wave reaches the 8
millisecond point (180º) the voltage will surge to 150 volts. This surge will hold
for 20 milliseconds before the voltage output returns to 120 volts. Trans-Time
is only selectable when the source is configured for AC output.
32.
Trans-Cycle
- gives the operator the flexibility to program whether the
transient voltage will occur continuously for each size wave of the test routine.
The operator has the choice of selecting 0 or 1-9999. Trans-Cycle is only
selectable when the source is configured for AC output.
33.
Phase Set (Option 08 only)
- gives the operator the flexibility to adjust
the hi- and lo-limits for all phases during 1 3W and 3 4W modes. Selecting
L1-N will allow the operator to set the limits for the instrument providing L1-N
output voltage. Selecting L2-N will allow the operator to set the limits for the
instrument providing the L2-N output voltage. Selecting L3-N will allow the
operator to set the limits for the instrument providing the L3-N output voltage.
Limits must be set for each phase in all steps (PROGRAM mode only).
34.
Prompt (PROGRAM Mode only)
- gives the operator the flexibility to
program a message unique to a particular step. The message will be shown
on the LCD graphic display prior to the test beginning for that particular step
test routine. At this point the test routine will be interrupted and the operator
must press the Test/Reset key to begin the test sequence.
24.
Connect (PROGRAM Mode only)
- gives the operator the flexibility to
program whether one step will be linked or connected to another step. For
example in order to links step one to step two, the Connect parameter must
be turned ON. Steps can only be connected in sequential order.
4.6 Editing Test Parameters
To edit testing parameters press the Edit soft key from the set screen. The
following screen will be displayed if the system parameters are set to PROGRAM
Mode: