background image

12

model S-254 UV optical sensor

The optional S-254 UV optical sensor measures the
relative output of ultraviolet
light within the UV treat-
ment chamber. Degradation
of the UV lamps, fouling of
the quartz sleeves, and
increased turbidity of the
water will affect the sensor
reading.

The UV meter assembly and
instrumentation setpoint
potentiometers are located on
the electrical enclosure.
The peripheral alarm
contacts are accessible
from the back side of the
meter. A shielded signal
cable connects the sensor
meter assembly to the sensor probe. The sensor probe
is located in the sensor port fitting on the UV treatment
chamber.

The meter must be reset after the first 100 hours of
operation on new UV lamps, both at the time of installa-
tion and each time the UV lamps are replaced.

SETTING THE UV METER

1. Gradually start the normal flow rate.

2. Bring the unit up to

normal operating
pressure.

3. Turn the UV lamps

on for a minimum of
15 minutes.

4. Open the S-254 meter

door.

5. Locate the blue potentiometer for setting the sensitiv-

ity “R3L” and blue potentiometer for setting the alarm
“R8L” located on the top right side of the PCB board.

6. Using a flat screwdriver, adjust the R3L sensitivity

potentiometer to 100%.

7. Using a flat screwdriver, adjust the R3L sensitivity

potentiometer to 50%.

ALARM CONTACTS (OPTIONAL)

Normally Closed and Normally Open alarm contacts are
also provided for use with user-supplied peripheral
equipment, such as remote alarms or solenoid valves.
Dry relay contacts are rated as follows: 0.52 AMP 120
VAC., 0.25 AMP 240 VAC

If your application requires higher contact ratings,
use a slave relay. Connection of peripheral equip-
ment is the responsibility of the user.
See wiring schematic in back pocket of this manual

SENSOR ALARM

The S-254 UV optical sensor provides information
about the relative amount of UV passing through the
water. The reading is affected by quartz sleeve fouling
and/or the germicidal lamp efficiency. The UV sensor is
sensitive to, and reflects changes in, UV transmission
when a substantial change occurs in the normal
operating flow rate, temperature, operating pressure, or
quality of the fluid flowing in the system.

When the UV intensity falls below the minimum
standard due to any of the above conditions, the sensor
alarm light (red LED) will come “on” and corrective
action must be taken to optimize the performance of the
UV treatment unit.

1. Examine your system for any significant changes in

normal operating conditions.

2. Verify that all UV lamps are operating. Verify the

quartz sleeves are clean.

If quartz sleeve fouling has caused the optical sensor
alarm to activate, the quartz sleeves must be cleaned.
At the same time, you must clean the sensor probe
quartz window. Please find this information in the
maintenance section of this manual.

more

S-254 optical sensor meter

8. Using a flat screwdriver, adjust the R8L alarm

potentiometer until the red LED barely comes on.

9. Using a flat screwdriver, set the R3L sensitivity

potentiometer back to 100%.

10. Turn the R3L sensitivity potentiometer to 50%; the red

LED should light.

11. Turn the R3L sensitivity potentiometer back to

100%.

R3L R8L

Summary of Contents for MP-2-SL

Page 1: ...s imperative that those responsible for the installation of this equipment as well as operating personnel read this manual and carefully follow all instructions and guidelines EQUIPMENT OPERATORS AND...

Page 2: ......

Page 3: ...eves 9 Installing the UV Lamps 10 Powering Up the Ultraviolet UV Unit 11 Monitoring Devices 11 UV Lamp Operational Display Panel 11 Running Time Meter 11 S 254 UV Optical Sensor 12 Setting the UV Mete...

Page 4: ...comply may result in severe or fatal electrical shock 6 Install the unit away from undue vibration that can damage the electrical components and UV lamps 7 Ensure all water connections flanges and co...

Page 5: ...bine the electrical components and the treatment chamber in one integral unit SL 10A This unit mounts vertically Inside the cabinet housing are the ballast and the UV treatment chamber The housing has...

Page 6: ...ed or repaired our option and returned to sender prepaid Before returning any part contact Aquafine Corporation for return authorization and shipping instructions This guarantee is void if the equipme...

Page 7: ...verheat problems described above where to install the unit Install the UV treatment unit in a horizontal position in a sheltered area with ample ventilation Ambient temperatures surrounding the unit s...

Page 8: ...e contact your local Aquafine represen tative or the factory directly wiring the unit A 1 2 conduit knockout 7 8 dia has been provided to attach a field installed junction box This box should be direc...

Page 9: ...8 plumbing schematic for all SL series ultraviolet treatment units MOUNTING BRACKET 2PLACES LED UV LAMP OPERATIONAL DISPLAY PANEL RUNNING TIME METER ModelSL 10A Service clearance required on this end...

Page 10: ...and o ring by placing the o ring into the internal relief of the compression nut below the threaded area The o ring should fit into the compression nut 9 Place the compression nut and o ring onto the...

Page 11: ...base into the rubber socket Push until you feel a firm bottomed out connection 8 Carefully slide the rubber boot portion of the lamp socket over the end of the lamp CAUTION DO NOT INSERT UV LAMPS WITH...

Page 12: ...hrough the system ensure there are no system leaks and no piping connection leaks All earth ground connections are properly made All lamp connections are properly made 1 Verify that all incoming power...

Page 13: ...lves Dry relay contacts are rated as follows 0 52 AMP 120 VAC 0 25 AMP 240 VAC If your application requires higher contact ratings use a slave relay Connection of peripheral equip ment is the responsi...

Page 14: ...empera turefallsto100 F 38 C HIGH TEMPERATURE HT UNITS The device automatically turns the UV lamps off when the water temperature reaches 170 F 77 C and auto matically turns the UV lamps back on when...

Page 15: ...r to one end of the lamp socket and the other end of the voltmeter of the opposite end of the lamp socket 7 Turn power on to the UV unit 8 Record the reading 9 Turn power off 10 Listed below please fi...

Page 16: ...less of the number of hours of operation due to normal operational degradation of the UV lamp CAUTION Prior to remov ing the cover to access the lamps and quartz sleeves you must release all pres sure...

Page 17: ...onnect ballast 9 Restore the power to the UV unit by turning the unit isolator switch to the ON position or by closing the upstream switch or breaker measuring performance Every UV treatment unit shou...

Page 18: ...or three and one half minutes reduce the flow to a reasonable stream of water not less than 50 of full flow Continue flowing to drain three additional minutes 4 Remove any temporary tubing used for fl...

Page 19: ...EPDM 4253 Viton 16458 Chemraz White 16600 Silicone 12967 2 LampSocket 16184 3 Running Time Meter 120 Volt 15843 4 Running Time Meter 240 Volt 15844 COMMON TO ALL SL 1 MP 2 SL APPLICATIONS 5 Quartz Sl...

Page 20: ...eduction Applications Size 15 Length 3052 14 ValidatedUVLampsDE185nmYellow TOC Reduction Applications Size 15 Length 16715 15 Quartz Sleeve 3150 16 Ballast 120Volt 4035 17 Ballast 240Volt 3493 18 Comp...

Reviews: