background image

16

more

cleaning the sensor

probe window

1. Shut off the flow and release the pressure.

2. Shut off all power to the UV unit.

3. Remove the coaxial cable by pushing and turning

the BNC terminal counterclockwise. The connector
will slide off.

4. Twist and pull the probe from the sensor fitting.

5. Use a lint-free cloth with alcohol and very carefully

wipe the lens face on the front of the probe. Failure
to do so may result in false readings.

6. Replace the probe and the coaxial cable.

7. Turn the power on and resume operation. Whenever

new lamps have been installed, follow instructions
on page 12 under “Setting the UV Meter.”

ballast replacement

Ballast replacement is not part of the UV unit’s routine
maintenance. However, in the event that a ballast needs
to be replaced, the following procedure should be
followed:

1. Power down the UV unit by turning the unit isolator

switch, if installed, to the “OFF” position. If not,
ensure that power to the unit is removed by opening
the switch or breaker upstream of the UV unit.

2. Locate the old ballast to be replaced. Refer to the

electrical diagram, if necessary.

3. Isolate the old ballast from rest of system by discon-

necting the ballast connector.

4. Using 5/32 L-shaped hex wrench completely

remove the bolt securing the ballast to the sub
panel. Put the bolt aside.

5. Remove the old ballast by pulling it towards you.

Discard the old or defective ballast.

6. Install the new ballast, securing it with the bolt

removed earlier.

7. Tighten the bolt with 5/32 L-shaped hex wrench.

8. Reconnect ballast.

9. Restore the power to the UV unit by turning the unit

isolator switch to the “ON” position or by closing
the upstream switch or breaker.

measuring performance

Every UV treatment unit should be tested periodically to
verify actual efficiency. Regardless of the intended
application or any optional equipment which may have
been provided with your UV unit, the most accurate and
dependable procedure is to conduct post-UV sample
analysis in accordance with standard testing methods.

Periodic sample collection and testing should be
scheduled as often as the user deems sufficient to be
assured the quality of the Aquafine ultraviolet unit
effluent is acceptable.

obtaining proper

water samples

Our experience has shown that the vast majority of
unsatisfactory post-UV bacteriological samples are
directly related to improper sample-taking techniques.

There are a  variety of commercial sample collection
apparatuses available, and should you choose one, be
sure to follow the manufacturer’s recommended
procedures.

1. Turn off the water to the unit.

2. Disconnect the electrical circuit.

3. Drain the UV treatment chamber.

4. Wear clean cotton gloves to prevent contami-

nation of the quartz sleeves and UV lamps.

5. Remove the  lamp sockets.

6. Remove the ultraviolet lamps from inside the

quartz sleeves.

7. Using the appropriate compression nut tool,

loosen the compression nuts and carefully
remove the quartz sleeves.

8. Wash the quartz sleeves with mild soap and hot

water and rinse with clean, hot water.

Should this be insufficient to clean the quartz
sleeves, they should be replaced. To place an order,
contact Aquafine or your local representative. To
replace the quartz sleeves, follow the procedures on
page 9.

Summary of Contents for MP-2-SL

Page 1: ...s imperative that those responsible for the installation of this equipment as well as operating personnel read this manual and carefully follow all instructions and guidelines EQUIPMENT OPERATORS AND...

Page 2: ......

Page 3: ...eves 9 Installing the UV Lamps 10 Powering Up the Ultraviolet UV Unit 11 Monitoring Devices 11 UV Lamp Operational Display Panel 11 Running Time Meter 11 S 254 UV Optical Sensor 12 Setting the UV Mete...

Page 4: ...comply may result in severe or fatal electrical shock 6 Install the unit away from undue vibration that can damage the electrical components and UV lamps 7 Ensure all water connections flanges and co...

Page 5: ...bine the electrical components and the treatment chamber in one integral unit SL 10A This unit mounts vertically Inside the cabinet housing are the ballast and the UV treatment chamber The housing has...

Page 6: ...ed or repaired our option and returned to sender prepaid Before returning any part contact Aquafine Corporation for return authorization and shipping instructions This guarantee is void if the equipme...

Page 7: ...verheat problems described above where to install the unit Install the UV treatment unit in a horizontal position in a sheltered area with ample ventilation Ambient temperatures surrounding the unit s...

Page 8: ...e contact your local Aquafine represen tative or the factory directly wiring the unit A 1 2 conduit knockout 7 8 dia has been provided to attach a field installed junction box This box should be direc...

Page 9: ...8 plumbing schematic for all SL series ultraviolet treatment units MOUNTING BRACKET 2PLACES LED UV LAMP OPERATIONAL DISPLAY PANEL RUNNING TIME METER ModelSL 10A Service clearance required on this end...

Page 10: ...and o ring by placing the o ring into the internal relief of the compression nut below the threaded area The o ring should fit into the compression nut 9 Place the compression nut and o ring onto the...

Page 11: ...base into the rubber socket Push until you feel a firm bottomed out connection 8 Carefully slide the rubber boot portion of the lamp socket over the end of the lamp CAUTION DO NOT INSERT UV LAMPS WITH...

Page 12: ...hrough the system ensure there are no system leaks and no piping connection leaks All earth ground connections are properly made All lamp connections are properly made 1 Verify that all incoming power...

Page 13: ...lves Dry relay contacts are rated as follows 0 52 AMP 120 VAC 0 25 AMP 240 VAC If your application requires higher contact ratings use a slave relay Connection of peripheral equip ment is the responsi...

Page 14: ...empera turefallsto100 F 38 C HIGH TEMPERATURE HT UNITS The device automatically turns the UV lamps off when the water temperature reaches 170 F 77 C and auto matically turns the UV lamps back on when...

Page 15: ...r to one end of the lamp socket and the other end of the voltmeter of the opposite end of the lamp socket 7 Turn power on to the UV unit 8 Record the reading 9 Turn power off 10 Listed below please fi...

Page 16: ...less of the number of hours of operation due to normal operational degradation of the UV lamp CAUTION Prior to remov ing the cover to access the lamps and quartz sleeves you must release all pres sure...

Page 17: ...onnect ballast 9 Restore the power to the UV unit by turning the unit isolator switch to the ON position or by closing the upstream switch or breaker measuring performance Every UV treatment unit shou...

Page 18: ...or three and one half minutes reduce the flow to a reasonable stream of water not less than 50 of full flow Continue flowing to drain three additional minutes 4 Remove any temporary tubing used for fl...

Page 19: ...EPDM 4253 Viton 16458 Chemraz White 16600 Silicone 12967 2 LampSocket 16184 3 Running Time Meter 120 Volt 15843 4 Running Time Meter 240 Volt 15844 COMMON TO ALL SL 1 MP 2 SL APPLICATIONS 5 Quartz Sl...

Page 20: ...eduction Applications Size 15 Length 3052 14 ValidatedUVLampsDE185nmYellow TOC Reduction Applications Size 15 Length 16715 15 Quartz Sleeve 3150 16 Ballast 120Volt 4035 17 Ballast 240Volt 3493 18 Comp...

Reviews:

Related manuals for MP-2-SL