LOG-AQUA

 

 

- 14 -

 

ANTENNA MOUNTING 
 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

DOs

 

 

Ensure  the  antenna  connector  is  
secured  tightly  to  the  panel 
connector. 

 

Mount antenna under the open sky or 
at  least  50cm  (20in.)  beneath  any 
object. 

 

Mount antenna at least 5 to 10cm (2 
to 4in.) away from wall. 

 

Mount  antenna  at  least  5cm  (2in.)  
away from the LOG-AQUA. 

 

 Finalize  the  antenna  setup  to  the 
exact  physical  conditions.  For 
example, close the lid, close the door, 
etc. 

 

Make  sure  you  have  a  signal  and  a  
successful  data  transmission  via  the 
user interface 

 

If needed during installation, use the 

Transmit  Now 

command  with  the 

magnetic  LOG-AQUA  Activator  to 
initiate more rapid transmissions

 

DONTs 

 

Do  not  attach  the  antenna  to  the  LOG-
AQUA. 

 

Do  not  wrap  cables,  zip  ties,  or  other 
items around the antenna. 
 

IMPORTANT NOTE :  

There are a few- minutes of delay between a 
successful transmission and the use of either 
of  these  methods  for  initiating  data 
transmission.  Repeating  the  use  of  either 
method will not expedite data transmission. 
 

WARNING :  

If  installing  in  a  highly  corrosive 
environment such as a sewer, apply technical 
grease  to  the  antenna  and  sensor  field 
attachable  connectors  after  securing  them 
to the panel connectors.  

 

Summary of Contents for LOG-AQUA

Page 1: ...LOG AQUA User Manual Version 1 1 ...

Page 2: ...e Quick Start Guide in a controlled environment prior to installation Set up activate and successfully test the entire system LOG AQUA sensors and antenna connection indoors in a controlled environment before going to the field for installation ...

Page 3: ...o 17 4 cm 6 9 po IP 68 NEMA 6P Wavelet mechanical enclosure Slots for zip ties and or bolts for mounting Panel connectors protector 13 3 cm 5 2 po Magnetic Activator FRONT BOTTOM External power port Serial Port Discrete port Cellular antenna port ...

Page 4: ...LOSURE INTERNAL LOWER ENCLOSURE Military grade lithium battery pack field replaceable non rechargeable Interface board ON OFF switch Indicator LEDs GSM antenna SIM card slot 2 Memory card GPS antenna SIM card slot 1 Cellular modem ...

Page 5: ...LOG AQUA 5 COMPONENTS 1 3 LOG AQUA Logger Activator 1x 1x Panel connectors protector 1 3 Cellular antenna and mounting bracket REQUIRED and sold separately 4x 2x 4x Zip ties screws and anchors ...

Page 6: ... on the LOG AQUA See page 17 for more details Turn the adjustable stainless steel end piece to secure the field attachable connector to the LOG AQUA CAUTION Do not turn the black plastic hood of the connector Turning the black hood may cause the wires to disconnect break and or damage the connector pins TURN HERE DO NOT TURN HERE ...

Page 7: ...e but it must remain connected after it is properly secured to the appropriate panel connectors Place the magnetic activator on the embossed logo on the front of the LOG AQUA enclosure The LOG AQUA will initiate a 15 minute test mode of sampling and send a few transmissions of data The device will then return to its default configuration Magnetic activator ...

Page 8: ...ector Arches a Insert the two lower clips into the two lower holes of the LOG AQUA enclosure b Snap the upper clips into place in the two grooves above the panel connector a b CAUTION The protector is provided to shield the connector from tampering or over exposure that could result in disconnected wiring If you need to remove the protector grasp it at the protector arches and pull up The protecto...

Page 9: ...VATION The LED light on the back cover of the Wavelet indicates the device status Access the AQUALABO user interface at XXXX by using your log in credentials Expect data to appear 5 10 minutes after the test mode is initiated ...

Page 10: ...operly then the test was successful You can now install the LOG AQUA in the field and begin to collect and view your data If data does not transmit properly change the LOG AQUA installation location and activate again If the problem persists contact the Aqualabo Team for help sav aqualabo fr 33 0 4 11 71 97 41 ...

Page 11: ...ing and sensor Wait for at least 15 minutes then log in to the user interface to confirm successful transmission UNDERGROUND INDOOR If the LOG AQUA is installed in an area with weak cellular signal activate the LOG AQUA and place in the intended installation location with closed doors access hatch Wait for at least 15 minutes then log in to the user interface to confirm the updated location on the...

Page 12: ...LOG AQUA 12 MONTAGE DU LOG AQUA Additional tools and or materials might be required not included Rotary hammer Screwdriver with bit ...

Page 13: ...LOG AQUA 13 LOG AQUA MOUNTING Secure the LOG AQUA on a wall pipe or other secure mount location by using zip ties or screws OU Wall mounted OU 4x 4x 1x 1x Pipe mounted 2x ...

Page 14: ... interface If needed during installation use the Transmit Now command with the magnetic LOG AQUA Activator to initiate more rapid transmissions DONTs Do not attach the antenna to the LOG AQUA Do not wrap cables zip ties or other items around the antenna IMPORTANT NOTE There are a few minutes of delay between a successful transmission and the use of either of these methods for initiating data trans...

Page 15: ...If the Wavelet does not transmit move the antenna to a different position If the Wavelet still does not transmit after multiple attempts at repositioning the antenna consider using an alternative solution such as an in road or a high gain antenna ...

Page 16: ...LOG AQUA 16 YOUR LOG AQUA IS SUCCESSFULLY INSTALLED ...

Page 17: ...LOG AQUA 17 LOG AQUA PINOUT Panel Connector Inputs 1 4x analog and 1x discrete 2 RS485 RS232 SDI 12 16 channels 3 4x discrete 4 6 24VDC 4 3 1 2 ...

Page 18: ...LOG AQUA 18 AQUALABO 90 Rue du Professeur Paul Milliez 94506 Champigny sur Marne France Tel 33 0 1 55 09 10 10 Hotline 33 0 4 11 71 97 41 E mail info aqualabo fr Website https www aqualabo fr ...

Reviews: