2
Aqua-Check Water Analyzer Operator’s Manual
Rev. 2.1
Specifications
Data Storage
•
The Water Analyzer stores 199 sets of readings. A log number is assigned
when the storage button is pressed and displayed when readings are being
reviewed.
Data Review
•
Stored readings can be reviewed in Review Log Mode.
Battery
•
6 V rechargeable sealed lead acid.
Discharge Time
•
8 hours (6 hours with RS-232 option)
•
Low battery indicator and automatic shutoff to prevent calibration and stored
reading memory loss
Sensor
Unit
Resolution
Range
Accuracy
Compensation
pH
pH
0.01
1–13
0.02
Temp (auto)
mV
1
±500
±2%
None
Temp
°C
.1°
0–50
1
None
°F
.1°
32–122
1.8
None
Cond
mhos
.1 µmho
0–99.9
±3%
Temp (auto)
1 µmho
0–999
±3%
(20° or 25°C)
.01 mmho
0–9.99
±3%
or none
.1 mmho
0–99.9
±3%
Siemens
.1 µS
0–99.9
±3%
Temp (auto)
1 µS
0–999
±3%
(20° or 25°C)
.01 mS
0–9.99
±3%
or none
.1 mS
0–99.9
±3%
TDS
.1 ppm
0–49.8
±3%
Temp (auto)
1 ppm
0–498
±3%
(20° or 25°C)
.01 ppt
0–4.98
±3%
or none
.1 ppt
0–49.8
±3%
Oxygen
ppm
.1 ppm
0–20.0
±2% to 5% Temp (auto),
Salinity, Opt.
%Sat
0
150.0 %Sat
±2% to 5% Temp (auto),
Salinity, Opt.
Press (auto)
Summary of Contents for Aqua-Check
Page 5: ...Chapter 5 Troubleshooting Troubleshooting 25 Index Index 27 ...
Page 25: ...20 Aqua Check Water Analyzer Operator s Manual Rev 2 1 Notes ...
Page 31: ...26 Aqua Check Water Analyzer Operator s Manual Rev 2 1 Notes ...
Page 36: ...Index 31 Notes ...
Page 37: ...4 30 Royal Crest Crt Markham On L3R 9W8 Tel 905 946 1064 FAX 905 946 8064 ...