Model 2500A225C
Rev A
13
CAUTION:
Make sure any external 7-16 DIN connector components that may be
attached to the amplifiers RF output connector are suitably rated for
power handling and frequency range of this product. Failure to use
properly rated mating components such as RF cables, adapters, or
directional couplers could cause serious injury to the user or amplifier at
the RF Output when the amplifier is under power.
2.4.3
IEEE-488, USB, and RS-232 Interfaces
The Model 2500A225C’s remote interfaces allow remote control, via a computer, of all amplifier functions
(except for the Keylock Switch position) that can be controlled from the front panel. When the amplifier is in
the Remote mode (as determined by the position of the Keylock Switch), a special Remote display is shown.
Front panel controls are deactivated, except for the Keylock Switch, which can be rotated to LOCAL or
INHIBIT to prevent remote control commands from being executed.
Refer to Section 2.10 for information on remote operation of the amplifier.
2.4.4
Safety Interlock Connector
Located on the left side of the rear panel, as viewed from the rear, are the remote interface connectors and safety
interlock connectors. The 15-pin D-sub miniature female safety interlock connector provides two separate
interlock loops. Pins 1 and 8 are used for situations where the amplifier can be left on but forced into
STANDBY mode. This interlock is called Inhibit Interlock. Pins 10 and 14 are used for situations that demand
the safest possible condition, with the AC power to all main circuits disconnected. This interlock is called AC
Interlock. Both loops must be closed, or jumpered, for normal operation.
2.4.5
RF Sample Ports
The Forward and Reverse power sample port connectors are Type N female and can be located on the front or
rear of the amplifier depending on ordered configuration. Cables and equipment connected to these ports should
exhibit resistive 50Ω impedance throughout the band. The sample ports are normally used for interfacing with
external power meters. A calibration table that provides a list of external RF power meter offset values is
provided. The offset values completely characterize the directional coupler forward and reverse sample port
circuit attenuation across the frequency band. This table is provided electronically as part of the test data packet
accompanying this manual.
Summary of Contents for 0053007
Page 2: ......
Page 4: ......
Page 6: ......
Page 18: ...Model 2500A225C iv ...
Page 28: ......
Page 70: ...Model 2500A225C 46 Rev A ...
Page 80: ...Model 2500A225C 56 Rev A Figure 4 7 SMA Module Input Figure 4 8 Final Module RF Output ...
Page 82: ...Model 2500A225C 58 Rev A ...
Page 84: ...Model 2500A225C 60 Rev A ...
Page 86: ......