20 www.arapharma.com
www.arapharma.com 21
Step 1
Insert a test strip into the meter’s test strip
port, with the contact bars facing upwards.
Gently push the test strip into the port until
the meter beeps. Be careful not to bend the
strip while pushing it in. The
symbol
will show up.
Control Solution Testing
• Whenever you open a new vial of test strips,
• If the meter or test strips do not function properly,
• If your symptoms are inconsistent with the blood glucose
test results and you feel that the meter or test strips are not
working properly,
• If you drop or damage the meter.
Step 2
Shake the Control Solution bottle
before each test. Remove the cap
and squeeze the bottle to discard
the first drop.
Then wipe the tip with a clean
tissue or cloth. Dispense a drop
of control solution onto a clean
non-absorbent surface. It helps to
squeeze a drop onto the top of the cap as shown. After the
symbol appears on the display, apply the solution to the
narrow edge of the test strip until the meter beeps. Make sure
the confirmation window fills completely.
Step 3
A test result will appear after the meter counts
down from 5 to 1.
After your control solution result appears on
the display, press for 3 seconds till the ‘check’
appears on the display. When the ‘check’ is
displayed, the result is not stored in the meter’s
memory and is not included in the 14-day
averages.
Note:
The meter may switch off if the control solution sample
is not applied within 2 minutes of the
symbol appearing
Step 4
Compare the result displayed on the meter
to the range printed on the test strip vial.
The result should fall within the range.
Used strips should be discarded safely in
appropriate containers.
Caution:
The range printed on the test strip vial is for the
First Canadian Health SPIRIT Control Solution only. It has
nothing to do with your blood glucose level.
on the screen. If the meter turns off, remove the strip, reinsert,
and start from step 1.