Chapter 4
Startup & Configuring
4.1
Startup
NOTE:
Apply only inlet voltages from 9 to 30 Vdc.
The Model 1095A/C is
protected from reversed polarity of the inlet voltage, however if reversed
the clock will not operate
.
The Model 1095A/C may be configured without being synchronized
to the GPS. However, the Unlocked and Fault LEDs will be lit and time
will not appear correctly on the Model 1095C. See page 25 for Model
1095C startup details.
1. Be sure to verify the inlet voltage level and polarity before con-
necting power. Verify connection by referring to Figure 2.2.
2.
Clocks with Internal Antenna:
Position the clock so that it can
receive GPS signals.
Sometimes being positioned against a
window will allow reception of GPS signals for synchronization.
For best GPS reception, follow the guidelines in Chapter 3.
Clocks with External Antenna:
For clocks using external GNSS
antennas, make sure that the antenna cable is connected to the
Type F connector at the base of the clock and position the antenna
outside so it can receive GPS signals.
3. Compare startup characteristics with description on the following
pages, which normally occurs when power is applied to the Model
1095A/C with normal GPS reception.