WAV - Waveform
To view or record signal waveforms on the 928A, select
one or both of the measurement channels, for voltage
and/or current. You can view up to two voltages, two
currents or one voltage and one current. It does not
matter which channel you select (A or B) to measure
voltage or current.
Setup
1. Make sure that Channel A and B are set up prop-
erly with at least one channel configured for volt-
age and/or current. If necessary, refer to Config-
uring Channels A and B on page 23.
2. Press the appropriate Signal Selection key in or-
der to measure the signal properly - for example,
press
2nd
>
AV
to select Channel A to measure
voltage.
3. Connect the probe(s) between the 928A and the
circuit under test.
4. Press
WAV
and read the Channel A graph on
the display.
Channel A
5. Press
WAV
again to view the channel B graph
on the display.
6. Press
WAV
a third time for a combined view of
channels A and B on the display.
42
Summary of Contents for 928A
Page 1: ...MODEL 928A USER S MANUAL Arbiter Systems Inc Paso Robles CA 93446 U S A ...
Page 6: ...vi ...
Page 12: ...xii ...
Page 14: ...xiv ...
Page 57: ...Channel B Combined Caution For current probe use see CT Probe Use on page 4 43 ...
Page 112: ...CT Cable Connector Figure 16 CT Cable Connector side view Figure 17 CT Connector End View 98 ...
Page 114: ...100 ...
Page 115: ...CE Mark Certification ...