Configuring Triggers
Note that you can only configure TRIG from the 928A
itself, not in Mlink
.
To configure TRIG, press
2nd
>
TRIG
>
2nd
>
MENU
. Once in this menu, step through each condi-
tion and view or edit any value. There are 8 individual
triggers available.
1. Press
ENT
and
or
to select trigger.
2. Use
ENT
>
and , or the number keys to the
configure triggering conditions.
3. Highlight
<
Store And Exit
>
and press
ENT
to
store new values.
4. Press
ESC
to exit the configuration screens.
Trigger Setup Value Definitions
Trigger:
8 possible profiles available
Signal:
Allowable input signal values: INACTIVE
(means the trigger is not armed), CH-A MAG, CH-B
MAG, WATTS, VA, VAR, PF and FREQ.
Logic:
4 conditionals: x
<
Limit; x
>
Limit;
|
x
|
>
Limit;
|
x-ref
|
>
Limit; x = input signal value
Limit:
Numerical value compared to signal: floating
point.
Ref:
A nominal signal value (e.g. 60 Hz) used to com-
pare to Limit: floating point
Dwell:
The persistence in alarm condition before trig-
ger becomes active, in 50 millisecond increments (e.g.
20 = 1 second): an integer.