27
Tensor
•
Finally save new value with button [M], thereby the next menu item is active automatically.
Defaults for the set values of the end positions must cover at least 20% of full scale,
otherwise the error "PE" - "Parameter Error" is reported.
Example: Range 1..5 V => Covering (5-1) V x 20% = 0.8 V
Lower end position 0.5 V => Upper end position min. 0.5 V+0.8 V =1.3 V
7.2.2.3 Actual value output for the end positions
The actual feedback must be set analogous to the set values for both end positions.
C
.
L
C
.
t
•
Set with DIP switches 1 and 2 how the actual value is used, see page 20th.
•
Select menu item „CL“ or „CR“ and confirm with button [M].
•
Set the value depending on the DIP position:
− DIP1/DIP2 – ON/OFF: 0…10 V
− DIP1/DIP2 – OFF/ON: 0... 20 mA
•
Finally save the new value with button [M], thereby the next menu item is active automatically.
Defaults for the actual limits of the end positions must cover at least 20% of full scale,
otherwise the error "PE" - "Parameter Error" is reported.
Example: Set range 4 mA => Covering 4 mV x 20% = 0.8 mA
Lower end position 0.1 mA => Upper end position min. 0.9 mA (=0.1+0.8)
7.2.2.4 Fault message output
The contact to the terminals 56 and 57 is closed when the controller is in operation.
The following events will trigger the opening of the signal contact:
•
I-ACT is in hand mode
•
>>
bE
: Motor Error
•
>>
SE
: Sensor Error
•
>>
tE
: Range Error
•
>>
CE
: Calibration Error
•
>>
PE
: Parameter Error
Detailed description of errors see chapter 8.2.
A
.
o
For activating the message function:
•
Select menu item „Ao“ and confirm with button [M].
•
Activate output with button [R] (1), set inactive with button [L] (0).
•
Finally save new settings with buttonutton [M].
0 mA
(0 V optional)
4 mA
(2 V optional)
20 mA
(5 or 10 V optional)
free values [mA]
([V] optional)
ADVICE
!
ADVICE
!