8 English
Ranger Blood/Fluid Warming
Arizant Healthcare Inc.
Instructions for Use
Preparation and setup of the Ranger warming unit
Attach the Ranger warming unit to the IV pole� Tighten the
1�
pole clamp securely�
Caution
: To prevent tipping, clamp the Ranger warming unit
to an IV pole with a minimum 14" (35�6 cm) radius wheelbase
and at a height no higher than 44" (112 cm)� Failure to do so
may result in damage to the product or catheter site trauma�
Slide the fluid warming cassette into the slot in the
2�
warming unit� The cassette can only fit into the device
one way�
Prime the disposable set� For more information about
3�
priming the set, refer to instructions provided with the
disposable sets�
Place the bubble trap in the holder�
4�
Plug the power cord into an appropriate outlet� Turn the
5�
unit ON� In a few seconds the alphanumeric display will
illuminate� It takes less than 2 minutes to warm up to the
41°C set point temperature�
Begin infusion� When infusion is completed, remove
6�
the disposable set and discard according to institutional
protocol�
44"
(112 cm)
14" (35.6 cm)