12 English
Ranger Blood/Fluid Warming
Arizant Healthcare Inc.
Maintenance and Storage
Cleaning the Ranger blood/fluid warming unit
Clean the Ranger unit on an as-needed basis
caution
Do not immerse the warming unit in cleaning or sterilizing solutions� The unit is not liquid
•
proof�
Do not clean the warming unit with solvents� Damage to the case, label, and internal components
•
may result�
to
clean
the
exterior
of
the
warming
unit
:
Disconnect the Ranger warming unit from the power source�
1�
Wipe the outside of the unit with warm, soapy water, nonabrasive cleaning solutions, dilute
2�
bleach, or cold sterilants� Do not use abrasive materials�
Wipe with a dry, soft cloth�
3�
to
clean
the
heating
Plates
:
The Ranger blood/fluid hardware cleaning tool is intended to clean both heating plates of the
warming unit� It is not necessary to disassemble the warming unit to use the tool�
caution
Do not insert metallic instruments in the warming unit, as damage to the heating plates may
•
occur�
Do not use abrasive materials or solutions to clean the heater plates�
•
Do not allow spills to dry inside the unit, as this may make it more difficult to clean the unit�
•
The cleaning tool provides only superficial cleaning, it does not disinfect or sterilize the interior
•
of the unit�