41
CLEANING AND DISINFECTION
Clean and disinfect with disinfection
unit
(39 steps)
Remove Visual Residue (step 1 to 6 of 39)
1. Raise Classic Line 2 bathtub to a convenient
working position.
2. Make sure all functions are turned off.
3. Open the drainage plug to empty any remaining
water.
4. Detach and place the footrest at the bottom of
Classic Line 2 bathtub.
5. Shower and rinse Classic Line 2 bathtub with
water to remove visible residue. Start from the top
and move downwards.
6. Shower and rinse the footrest at the bottom of
Classic Line 2 bathtub with water to remove
visible residue. Turn over the footrest and rinse
again. See Figure 66.
Clean (step 7 to 24 of 39)
7. Put
on
protective gloves, protective glasses
and protective appron
. See Figure 67.
8. Remove the covers and take the disinfection
spray handle from the compartment.
See Figure 68.
9. Check the level of disinfectant.
80
70
50
40
30
Figure 66
Figure 67
Figure 68
Continue with the steps on the next page