EASYTRACK System Requirements
6
Ceiling Requirements:
The EASYTRACK must be installed to a rigid
ceiling which have a framework above it.
• Must have a maximum of 610 mm (24”)
between the trusses or joists in the
ceiling.
• Ceiling must be made of wood or
12.7 mm (1/2”) sheet rock or thicker.
Fig. 3
Joists or
Trusses
610 mm (24’’)
MAX.
12.7 mm (1/2’’)
Sheet rock
Joists or
Trusses
Sheet roc
k or tiles
Suspended Wires or Rods
• Ceiling must be clear of any dust,
grease, water or any foreign substance.
WARNING
: A build-up of nicotine, grease or
water on the ceiling affects the integrity of
the EASYTRACK System which could lead
in patient fall.
Floor Requirements:
• Flooring must comply to local building
code standards.
WARNING
: Termites or water damages may
affect the structure of the floor.
• Floor must be clear of any dust, grease,
water or any foreign substance.
• Floor must have a slope of less than
0.5°. See drawing below to determine
the floor slope. Use a level or contact a
contractor if unsure of floor slope.
Fig. 4
X
Y
X
Y
600 mm (24”)
5.2 mm (3/16”)
900 mm (36”)
7.9 mm (5/16”)
1200 mm (48”)
10.5 mm (7/16”)
Other Requirements:
WARNING
: Use the clip-on level included
with kit to ensure all posts are straight –
failure to do so may lead to injury.
• A maintenance check should be done
on a monthly basis to ensure all posts
are levelled, see “Care and
Maintenance” section.
• Make sure the padded surface of the
top plate and foot plate are kept clean
and are intact.
EASYTRACK
System Requirements