14

Cleaning and Disinfection

WARNING

To prevent cross-contamination, always 
follow the disinfection instructions in this 

IFU

.

WARNING

To avoid material damage and injury, clean 
and disinfect according to this 

IFU

.

•  No other chemicals are allowed.
•  Never clean with chlorine. 
•  Chlorine will deteriorate the surface of 

the material. 

WARNING

To avoid injury, always remove the sling 
before disinfecting the lift.

WARNING

To avoid injury, never wash a Flites. A Flites 
is only intended for “single patient use”.

All 

Passive Loop 

slings, except the Flites, should 

be cleaned when they are soiled and stained and 
between patients. 

Cleaning Instruction (7 steps)

(NOT Flites) 

1.  Detach the sling from the lift.
2.  Remove the stiffeners from the stiffener pockets, 

if any.

3.  Close all buckles and hook and loop attachment 

on the slings and belts prior to washing.

4.  Check the symbols on the product label for 

cleaning recommendations.

5.  Machine wash the sling at 70°C (158°F). In order 

to obtain suf

¿

 cient level of disinfection, please 

wash according to local hygiene legislation. 
If allowed according to the product label, tumble 
dry in low temperature, max 60°C (140°F). 

6. 

Do NOT

•  wash with rough surfaces or sharp objects
• steam
•  use any mechanical pressure, pressing or 

rolling

• use 

bleach

•  use gas sterilization
• use 

autoclave

• dry 

clean

• ironing

7.  Place the stiffeners back into the stiffener 

pockets, if any, before use.

Flites – Do NOT wash

The Flites is for single patient use. 
Do NOT clean, wash, disinfect, wipe or sterilize the 
Flites. If the Flites has been subjected to any such 
treatment, it shall be discarded.
The Flites is marked with the “Do NOT wash” symbol. 

(See Fig. 19 )

If the Flites is washed, the “Do NOT use/lift” symbol 
will appear. 

(See Fig. 20 )

Cleaning Chemicals

For all slings (NOT Flites) use common commercial 
detergent without optical brightener.
No other chemicals are allowed e.g. chlorine, 
softener, iodine based disinfectants, bromine and 
ozone.

Disinfection

The only allowed disinfection is disinfection by 
washing.
This does not apply to the Flites. Do NOT clean, 
wash or disinfect the Flites.

Fig. 19  

Fig. 20 

EN

Summary of Contents for MLAAS2000

Page 1: ...INSTRUCTIONS FOR USE Passive Loop Slings 04 SL 00 7ENAU 08 2019...

Page 2: ...tions may cause damage to all or parts of the system or equipment NOTE Means This is important information for the correct use of this system or equipment Contents Foreword 2 Customer Support 2 De nit...

Page 3: ...mself herself at all Cannot stand unsupported and is not able to bear weight not even partially Dependent on caregiver in most situations Physically demanding for caregiver Stimulation of remaining ab...

Page 4: ...ed over a small surface area when compared to lying in bed An individualised skin and holistic assessment of the patient should be undertaken before deciding on whether a sling should be left under a...

Page 5: ...Shoulder strap with loops 5 Leg strap with loop 6 Leg ap 7 Body section 8 Pulling handle 9 Head section 10 Flites Do NOT wash use lift label located on the outside of the sling 11 Padding 12 Leg stif...

Page 6: ...nfection instructions in this IFU 4 Check that the sling is clean If the sling is not clean see Cleaning and Disinfection on page 14 5 Always check the washing symbol on the Flites label If the Flites...

Page 7: ...Head Support intended to be seated on S M L XL MLA3000 272 kg 600 lb Standard padded low back quick t S M L XL MFA2100 272 kg 600 lb Flites with head support Disposable and Patient Speci c S M L XL X...

Page 8: ...MLA2000 272 kg 600 lb S M S M L XL S M L XL M L XL S M L XL MLA2000A 272 kg 600 lb M M L XL M L XL M L XL M L XL MLAAS2000 190 kg 418 lb S M S M L XL S M L XL M L XL S M L XL MLA4060 272 kg 600 lb S M...

Page 9: ...next page Loop Attachment Detachment 2 point spreader bar First attach the shoulder loops then attach the leg loops 4 point spreader bar First attach the shoulder loops in the front hooks Then attach...

Page 10: ...according to the lift IFU In Chair Wheelchair 20 steps 1 Place the lift nearby 2 Apply the brakes on the wheelchair 3 Check that the stiffeners are completely inside the stiffener pockets if any 4 St...

Page 11: ...nue to step 18 9 Patient not able to sit up Fold the sling lengthwise along the centre line outside facing outside The sling label is located on the outside 10 Log roll the patient Use local technique...

Page 12: ...ess 30 Make sure that All loops are securely attached All straps are straight not twisted The patient lays comfortably in the sling 31 If an adjustment is needed lower the patient and make sure that t...

Page 13: ...Fig 18 Removing the Sling In Bed 13 steps 1 Position the patient over the bed See Fig 16 2 Apply bed brakes and adjust the height of the bed to an ergonomic position if possible WARNING To avoid inju...

Page 14: ...158 F In order to obtain suf cient level of disinfection please wash according to local hygiene legislation If allowed according to the product label tumble dry in low temperature max 60 C 140 F 6 Do...

Page 15: ...ort is intact If damaged or missing and the underlying red Do NOT use lift symbol appears replace the sling immediately See Label on the Sling on page 18 When Soiled Stained and Between Patients Clean...

Page 16: ...patient s neck back outside of the sling Raise the back rest of the bed if available Adjust the spreader bar to a more reclined position Pay attention to the patient s head There are dif culties appl...

Page 17: ...0 C 68 F Operating and Storage 10 95 at 20 C 68 F Transport Recycling The device should be recycled according to national regulations Package The bag consists of soft plastic recyclable to local legis...

Page 18: ...G Only on Flites Disposable and patient speci c Do NOT wash symbol Located on the outside of the sling Do NOT use lift symbol The Flites has been washed Located on the outside of the sling Symbol for...

Page 19: ...9 0 6 87426222 E mail Italy promo arjo com MIDDLE EAST Arjo Middle East FZ LLC Of ce 908 9th Floor HQ Building North Tower Dubai Science Park Al Barsha South P O Box 11488 Dubai United Arab Emirates D...

Page 20: ...rgonomic patient handling personal hygiene disinfection diagnostics and the effective prevention of pressure ulcers and venous thromboembolism we help professionals across care environments to continu...

Reviews: