Care and Maintenance
23
Care and Maintenance
Lift Cleaning and Care
Note
: It is recommended that your
MAXI LITE
and its accessories are cleaned and
disinfected between each patient use, or
daily as a minimum. If the lift and equipment
needs cleaning, or is suspected of being
contaminated, follow the cleaning and
disinfection procedures recommended
below, before re-using the equipment.
For cleaning your lift and its accessories wipe
down with a damp cloth using warm water to
which a disinfectant / cleaner has been added
(e.g. “ArjoHuntleigh” - disinfectant / cleaner or
equivalent).
CAUTION
: Never use a wet cloth on the
control box, the battery support or the battery
pack as it may cause corrosion on electrical
components.
If a hot air dryer is used to dry the lift, the
temperature must not exceed 80°C (176°F.)
Do not use petroleum based solvents or
similar, as this may damage plastic parts.
For disinfection of contaminated lifts and their
accessories, use the preferred method of wiping
the product completely with “hard surface
disinfectant wipes” that are supplied
impregnated with a 70% v / v solution of Isopropyl
Alcohol.
A rubbing action will be necessary when using
the wipes to promote effective disinfection of
the surfaces.
When cleaning, pay special attention to parts
that are most likely to be touched such as:
• the handset;
• the control box;
• the battery pack;
• the lift handle;
• the spreader bar.