Service Procedures
16
0
0
1
.1
60
1
0.
3
3 re
v
. 8
Battery Charger Replacement
Required tools:
• None
Battery Charger 700.13510.33 (OBSOLETE)
This charger has been replaced by the wall
mounted charger 700-15567 which requires to
have the touch pad membrane 001-08100-33
(featuring a charge indicator light) on the unit.
For this indicator to function properly, the main
circuit board shall be replaced with one of
revision 13 or higher if not already present.
Battery Charger 700-24201
(OBSOLETE)
This charger has been replaced by the wall
mounted charger 700-15567 which requires to
have the touch pad membrane 001-08100-33
(featuring a charge indicator light) on the unit.
Caution:
Do not use charger 700-15567or 700-24201
with a PCB of revision below 13.
1. Before replacing the charger, remove the Main Plastic cover and check if the PCB is featuring a
5-pin connector (revision 13 or greater). If not, the PCB has to be replaced as well, refer to
“Main Circuit Board Replacement” on page 13 to replace the circuit board.
2. Refer to "Touch Pad Membrane Replacement” on page 15 to replace the touch pad membrane.
Limit Switch Replacement
Required tools:
•
3/16” Socket
1. Refer to "Main Plastic Cover Replacement” on page 9 to remove plastic covers.
2. Unplug the two wires from the limit switch.
3. Unscrew the two locknuts from the other side of the frame
with a 3/16” socket.
4. Install the new limit switch on the frame and screw in the
locknuts. Do not over tighten the locknuts. Make sure to
reposition the contact blades under the upper limit shaft as
shown.
Plug the two wires back into both ends of the limit switch. Never
use the center pin (see “Wiring Diagram” on page 34).
5. Refer to "Main Plastic Cover Replacement” on page 9 to
reinstall the plastic covers.
Upper
limit
shaft