Comparing the Control Solution Test Results
The test result of each control solution should be within the
range printed on the label of the test strip vial. Repeat the
control solution test if the test result falls outside of the range.
Out of range test results may occur in following situations:
If results continue to fall outside the range printed on the test
strip vial, the GLUCOCARD Shine Blood Glucose Test Strip and
GLUCOCARD Shine Connex Meter may not be working properly.
Do not
use your system and contact ARKRAY Technical
Customer Service: 800.566.8558, 24 hours a day, 7 days a week.
Situations
Do This
• When the control solution bottle was not
shaken well.
• When the meter, test strip or control
solution have been exposed to high or
low temperatures.
• When the first 1 or 2 drops of control
solution were not discarded or the tip of
the bottle was not wiped clean.
• When the meter is not functioning properly.
Repeat the
control solution
test by referring
to page 29–31.
• When the control solution is past the
expiration date printed on the bottle.
• When the control solution is past its
discard date (the date the bottle was
opened plus three (3) months).
• When the control solution is contaminated.
Discard the used
control solution
and repeat the
test using a new
bottle of control
solution.