Performance Characteristics
The performance of GLUCOCARD Shine Connex Blood Glucose
Monitoring System has been evaluated in laboratory and in
clinical tests.
Accuracy
The accuracy of the GLUCOCARD Shine Connex Blood Glucose
Monitoring System was assessed by comparing blood glucose
results obtained by patients with those obtained using a YSI
Model 2300 Glucose Analyzer, a laboratory instrument.
The following results were obtained by diabetic patients at
clinic centers.
Accuracy results for glucose concentration < 75 mg/dL
Accuracy results for glucose concentration ≥ 75 mg/dL
Slope
1.0223
Y-Intercept
-1.3686
Elation coefficient (r)
0.9934
Number of subjects
371
Range tested
48 mg/dL– 553 mg/dL
Within ± 5 mg/dL Within ± 10 mg/dL Within ± 15 mg/dL
61.0%
(25/41)
97.6%
(40/41)
100%
(41/41)
Within ± 5% Within ± 10% Within ± 15% Within ± 20%
70.0%
(231/330)
96.1%
(317/330)
100%
(330/330)
100%
(330/330)