5400 Series
TECSource
User’s Manual · Page 45
was 25.0°C, and
T Error
was set to 0.1°C/V, then the analog output
voltage would be 0.4V.
Temperature
– similar to the
T Error
function, but with this function, the
analog output voltage is proportional to the actual temperature, not
temperature error. When set to this function, a second menu item,
T
Range
, will appear to control the transfer function of temperature to
output voltage.
T Range
is in °C/V. The output function looks like this:
For
example
,
if
the
actual
temperature
was
25
.
4°C
and
T Range
was set
to 10°C/V, then the analog output voltage would be 2.54V.
ITE
– with this function, the analog output voltage is proportional to the
ITE current, with the current scaled such that at the current limit, 5V (or
-5V) is generated. The output function looks like this:
∗ 5
For
example
,
if
the
current limit (
I Lim
) was set to 15A, and the actual
current was -4.50A, the analog output voltage would be -1.5V.
Summary of Contents for TECSource 5400 Series