5.2 Troubleshooting
Symptom
Possible Cause
Corrective Action
Settings do not
appear on surgical
monitor
•
Pump/shaver/Apollo/Insufflator
not powered on
•
Pump/shaver/Apollo/Insufflator
display not enabled
•
Surgeon preferences set to not
display settings
•
Ensure
pump/shaver/Apollo/Insufflator are
powered on
•
Confirm
pump/shaver/Apollo/Insufflator
enabled in settings
•
Set surgeon preference to display
settings
Settings appear as
0 on monitor
•
Pump/shaver/Apollo/Insufflator
settings are set to 0
•
Integration cable is disconnected
•
Pump/shaver not powered on
•
Confirm
pump/shaver/Apollo/Insufflator
settings are not 0
•
Confirm integration cables are
connected
•
Ensure
pump/shaver/Apollo/Insufflator
powered on
Settings disappear
from surgical
monitor
•
Surgeon preferences set to
display on only settings changes
•
Set surgeon preference to always
display settings