DrillSaw Highspeed 200™
User’s Guide
DFU-0225-2 Rev. 0
Page 24 of 58
The motor connector plug cannot be properly inserted into the console
receptacle if the corresponding red dots are not aligned.
The foot control unit connector plug cannot be properly inserted into the
console receptacle if the corresponding black triangle of the console is not
aligned with the black arrows of the cable plug.
W A R N I N G !
Use
ONLY
Arthrex approved accessories. Other accessories may result in
increased emissions or decreased immunity of the system. Contact your Arthrex
representative for a complete list of accessories.
DO NOT
modify any accessory.
Failure to comply may result in injury to the patient and/or operating room staff.
W A R N I N G !
Do not
use in the presence of flammable anesthetics or oxidizing gases such as
nitrous oxide, oxygen, or endogenous gases. All oxygen connections must be
leak free for the duration of the surgical procedure.
W A R N I N G !
Before using the AR-200 for the first time, store it at room temperature for 24
hours.
4.6
Motor Detection
The AR-200C detects when a motor with cable has been activated. The Motor
Detection feature operates differently, depending on if an AR-200M or an
AR-200M-ISO motor is connected to the AR-200C console.