500-003111-000
5-8
5.5 CALIBRATION
Before delivery from the factory, the VMIVME-3111 board is fully
calibrated to ±10 V operation and conforms to all applicable specifications as listed
in Section 2. However, should recalibration be required, refer to Sections 5.5.1
through 5.5.3, and perform all of the calibration operations in the sequence shown.
The locations of all adjustments and test points are shown in Figure 5.5-1.
As delivered from the factory, all calibration adjustments are sealed
against accidental movement. However, the seals are easily broken for
recalibration. All adjustments should be resealed with a suitable fast-curing
sealing compound after recalibration has been completed.
5.5.1 Equipment
Required
a. Digital Voltmeter (DVM)
±1.000 VDC and ±10.000 VDC ranges;
five or more digits; ±0.005 percent of
voltage range
measurement accuracy;
10 M
Ω
minimum input impedance.
b. Digital Voltage Source
(Voltage Input Option Only)
±1.000 VDC and ±10.000 VDC output
voltage ranges; ±0.005 percent of range
accuracy. Maximum output impedance
10
Ω
.
c. Digital Current Source
(Current Input Option Only)
±0.040 A output current range; ±0.005
percent of range accuracy.
d. Chassis
VMEbus backplane or equivalent, J1
and J2 connectors, 68000 series master
controller, +5 ±0.1 VDC, 3 A power
supply.
e. Extender Board
VMEbus extender board.