31
Analog Inputs
1
Calibration for the current option is similar to regular calibration as described in the
Calibration Mode section on page 20 and the Target Calibration Voltage Register (TVC)
section, but with one difference. The value entered in the TCVH and TCVL must be
expressed as a voltage. Therefore, to find the input voltage applied, multiply the
calibration current times 400 and use this value in the equation in the Target Calibration
Voltage Register (TVC) section.