56
3
VMIVME-3126A High Resolution, Isolated Analog-to-Digital Converter Board
Target Calibration Voltage Register (TVC)
The Target Calibration Voltage (TCV) register consists of two registers. The user writes the
digital value of the expected result for a given calibration input voltage. The Target
Calibration Voltage HI Register contains the upper word of a 32-bit value. The Target
Calibration Voltage LO Register contains the lower word of a 32-bit value. Table 3-9 below
shows the HI register and Table 3-10 on page 57 shows the LO register. The gain and offset
coefficients are determined using the values in these registers and the data from the ADC.
A precision low noise voltage source should be used for the calibration process to ensure
the accuracy of the final corrected data.
Powerup/Reset Default = $0000
Table 3-9 Target Calibration Voltage Register HI’s Bit Map
Target Calibration Voltage Register HI (Offset $000A) Read/Write, Byte/Word
Bit 15
Bit 14
Bit 13
Bit 12
Bit 11
Bit 10
Bit 09
Bit 08
TVH15
TVH14
TVH13
TVH12
TVH11
TVH10
TVH9
TVH8
Bit 07
Bit 06
Bit 05
Bit 04
Bit 03
Bit 02
Bit 01
Bit 00
TVH7
TVH6
TVH5
TVH4
TVH3
TVH2
TVH1
TVH0