Channel PWR (Power Meter)
The Channel PWR Module turns the AE20401 into a Power
Meter capable of measuring the power of signals from DC to
500 MHz / 10 MHz to 8 GHz in the range of -55 dBm to +30 /
-5 dBm. This range may be further extended with external
attenuators whose value can be entered and included in the
displayed measurement.
Instead of the modes on Channel A and B, pressing
Mode
changes the display from dBm to mW, Vrms, Vp or Vpp. The
relationship between the units is as follows:
P
mW
=
10
(
P
dBm
10
)
V
rms
=
√
(
P
mW
∗
50
1000
)
V
p
=
V
rms
∗
√
2
V
pp
=
V
p
∗
2
Please pay close attention not to exceed the
maximum input power! Even short overloads
may cause damage to the input!